510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0697-2022Class II

Microbiologics QC Sets and Panels: Helix Elite. Extended Enteric Bacterial Verification Panel (Inactivated...

Microbiologics, Inc. / initiated 2017-06-06 / Terminated / root cause: manufacturing process

Microbiologics Inc began a recall of Microbiologics QC Sets and Panels: Helix Elite. Extended Enteric Bacterial Verification Panel (Inactivated Pellet) includes 21 pellets each comprised of the... on . FDA classified it as Class II and gives the reason as "Real-time shelf life testing failed at 24 months". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0697-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Microbiologics, Inc.
Product
Microbiologics QC Sets and Panels: Helix Elite. Extended Enteric Bacterial Verification Panel (Inactivated Pellet) includes 21 pellets each comprised of the following 8 organisms: 1. S. Typhimurium derived from NCTC 74, 2. S. sonnei derived from NCTC 12984, 3. E. coli derived from CDC 2010C-3114 (STEC), 4. C. jejuni derived from NCTC 11322, 5. P. shigelloides derived from NCTC 10363, 6. Y. enterocolitica derived from NCTC 11174, 7. E. coli derived from NCTC 11602 (ETEC), 8. V. parahaemolyticus derived from NCTC 13209
Product code
OHQ Multi-Analyte Controls Unassayed
Reason for recall
Real-time shelf life testing failed at 24 months
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Microbiologics notified customers on about 06/06/2017 via email containing a letter. The letters indicated the affected products and instructed customers to update their records with the corrected expiration date, which was provided for each affected lot.
Quantity in commerce
8 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, CA, DC, FL, IA, ID, KS, LA, MD, MN, MO, NE, NH, NJ, NY, OH, OK, PA, SC, TX, VA, WA, WV and the countries of Australia, Belgium, Canada, India, United Kingdom.
Product codes and lots
Catalog Number: 8191, Lot Number: 8191-01, UDI: 70845357041240

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0697-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0697-2022Class IITerminatedVoluntary: Firm initiated
Z-0698-2022Class IITerminatedVoluntary: Firm initiated
Z-0699-2022Class IITerminatedVoluntary: Firm initiated
Z-0700-2022Class IITerminatedVoluntary: Firm initiated
Z-0701-2022Class IITerminatedVoluntary: Firm initiated