Z-0788-2022Class III
LYFO Disk
Microbiologics, Inc. / initiated 2015-07-17 / Completed / root cause: labeling and UDI
Microbiologics Inc began a recall of LYFO Disk on . FDA classified it as Class III and gives the reason as "LYFO Disk Catalog #0983L is to contain Candida tropicalis. However, lot number 983-40-3 under Catalog #0983L, labeled as Candida tropicalis, is actually Yersinia entercolitica. This labeling error may cause a user's...". As of , the recall status on record is Completed, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0788-2022
- Classification
- Class III
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- Microbiologics, Inc.
- Product
- LYFO Disk
- Product code
- JTR Kit, Quality Control For Culture Media
- Reason for recall
- LYFO Disk Catalog #0983L is to contain Candida tropicalis. However, lot number 983-40-3 under Catalog #0983L, labeled as Candida tropicalis, is actually Yersinia entercolitica. This labeling error may cause a user's quality control to fail, resulting in testing needing to be repeated and delayed test results.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Microbiologics notified customers on 07/17/2015 via letter titled "Medical Device Recall". Instructions included to inspect inventory for affected products. If customers have received the product identified within the recall (LYFO DISK 0983L, Lot# 983-40-3), it is recommended that the customer destroys the product or return it to the recalling firm. Acknowledgement of the receipt of the recall letter was requested by completing a form and returning to Microbiologics. Questions or concerns can be directed to the Quality Systems Manager at 320.229.7050.
- Quantity in commerce
- 3 units
- Distribution
- Distribution is nationwide with three consignees located in Colorado, Georgia, and Ohio.
- Product codes and lots
- Catalog Number 0983L, UDI: 10845357023930, Lot Number (Expiration Date): 983-40-3 (02/11/2016)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0788-2022 | Class III | Completed | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY