Z-0664-2022Class II

REF: 5921-018-135, Stryker Color Cuff, 18" (Red), Quick Connect Single Port, Single Bladder

Stryker Sustainability Solutions / initiated 2022-01-21 / Terminated / root cause: labeling and UDI

Stryker Sustainability Solutions began a recall of REF: 5921-018-135, Stryker Color Cuff, 18" (Red), Quick Connect Single Port, Single Bladder on . FDA classified it as Class II and gives the reason as "Mislabeling; dual-bladder tourniquet cuffs are labeled as single-bladder inflatable cuffs". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0664-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Sustainability Solutions
Product
REF: 5921-018-135, Stryker Color Cuff, 18" (Red), Quick Connect Single Port, Single Bladder
Product code
KCY Tourniquet, Pneumatic
Reason for recall
Mislabeling; dual-bladder tourniquet cuffs are labeled as single-bladder inflatable cuffs.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
On February 4, 2022, Stryker issued an "Urgent: Medical Device Recall notification" to all affected customers. In addition to informing consignees about the recalled product, the firm ask customers to take the following actions: 1.) Please check your inventory for the devices from the affected lot and remove them from their point of use. 2.) Please complete the Urgent Medical Device Recall Business Reply Form (page 3) and indicate if any of the affected product remains in your inventory. Please return this completed form through one of the following methods: a. Email: [email protected] b. To your Stryker s Sustainability Solutions Sales Representative c. Fax: 480-763-5345 This form must be completed and returned to Stryker even if no affected product is found. 3.) Maintain awareness of this communication internally until all required actions have been completed within your facility. If any of the affected products have been forwarded to additional facilities (including but not limited to distribution warehouses), please contact these facilities and communicate this recall to them. Please instruct these facilities to complete the Business Reply Form (page 3) and return the completed form either via email to [email protected], to their Stryker s Sustainability Solutions sales representative, or via fax to 480-763-5345. Alternatively, provide contact details so Stryker can inform the recipients appropriately. If the form indicates that any affected product remains in your inventory, a prepaid shipping label will be issued for the return of the product. Credit will be issued to your facility for all devices that are returned prior to May 1st, 2022. 4.) Adverse reactions or quality problems experienced with the use of this product should be reported to: " Stryker s Sustainability Solutions Complaint Hotline: +1-888-763-8803 " http://www.stryker.com/productexperience " The FDA s MedWatch Adverse Event Reporting program either online, by
Quantity in commerce
40 units
Distribution
US: KS OK TX OUS: None
Product codes and lots
Lot # 12983349

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0664-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0664-2022Class IITerminatedVoluntary: Firm initiated