Z-0791-2022Class II
Lumbar Puncture Tray Pediatric 21G 2.5
Bard Peripheral Vascular Inc / initiated 2022-01-20 / Open, Classified / root cause: manufacturing process
Bard Peripheral Vascular Inc began a recall of LUMBAR PUNCTURE TRAY PEDIATRIC 21G 2.5 on . FDA classified it as Class II and gives the reason as "Due to lumbar puncture tray manometer failure mode of leaking and mating issues". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0791-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Bard Peripheral Vascular Inc
- Product
- Lumbar Puncture Tray Pediatric 21G 2.5
- Product code
- PWY Lumbar Puncture Tray (Adult & Pediatric)
- Reason for recall
- Due to lumbar puncture tray manometer failure mode of leaking and mating issues.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Between 01/28 and 02/01/2022, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter via email and FedEx 2-day mailings to customers informing them that between 07/02 to 12/15/2021 the firm received 59 customer complaints for lumbar puncture tray manometers with reported failure modes of leaking and mating issues. Customers are asked to take the following actions: 1. Check all inventory locations within their institution for the product listed in Table A and immediately destroy all affected product remaining in their possession. The product should have the sterile barrier broken and destroyed per their local facility s destruction process. 2. Share this notification with all users within their facilities to ensure they are also aware of this Urgent Medical Device Correction. 3. If Customers purchased this product from a distributor, contact their distributor for further instructions and credit resolution. 4. Complete and return the attached Customer Response Form to the Recalling Firm contact noted on the form confirming acknowledgement of the correction notification, whether or not they have any affected product, so that the Recalling Firm may acknowledge their receipt of this notification and process their credit. 5. Indicate on the response form the quantity of affected product identified at their facility and confirm that this inventory was destroyed. 6. Contact their Recalling Firm's representative to assist in this process or they may also contact the listed in the notification for further assistance. For further assistance contact: North American Regional Complaint Center - 1-844-8BD- LIFE (1-844-823-5433), Say Recall when prompted M-F 8am - 5pm CT
- Quantity in commerce
- 11,970 devices
- Distribution
- U.S. Nationwide Distribution: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S. (foreign) including states: Costa Rica and Hong Kong.
- Product codes and lots
- Catalog Number: 4304C Lot Numbers/UDI (GTIN, DI + PI): 0001414261 (01)10885403057052(17)220331(10)0001414261 0001414362 (01)10885403057052(17)220430(10)0001414362 0001415851 (01)10885403057052(17)220430(10)0001415851 0001415026 (01)10885403057052(17)220430(10)0001415026 0001421686 (01)10885403057052(17)220630(10)0001421686 0001424759 (01)10885403057052(17)220630(10)0001424759 0001426324 (01)10885403057052(17)220630(10)0001426324 0001428876 (01)10885403057052(17)220630(10)0001428876 0001432173 (01)10885403057052(17)220630(10)0001432173 0001434168 (01)10885403057052(17)220930(10)0001434168 0001438149 (01)10885403057052(17)221031(10)0001438149 0001438930 (01)10885403057052(17)221031(10)0001438930 0001441010 (01)10885403057052(17)221031(10)0001441010
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0790-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0791-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0792-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0793-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0794-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0795-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0796-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0797-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0798-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0799-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0800-2022 | Class II | Ongoing | Voluntary: Firm initiated |