Codman Cranial Access Kits- indicated when a craniotomy is required for placement of an intracranial pressure...
Integra LifeSciences Corp. began a recall of Codman Cranial Access Kits- indicated when a craniotomy is required for placement of an intracranial pressure (ICP) monitoring device and/or cerebrospinal... on . FDA classified it as Class II and gives the reason as "Incorrect extended expiration dates identified on the outer box label only, if product is used past the correct expiration date there is a potential for pain, infection, or adverse tissue reaction to occur". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0818-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Integra LifeSciences Corporation
- Product
- Codman Cranial Access Kits- indicated when a craniotomy is required for placement of an intracranial pressure (ICP) monitoring device and/or cerebrospinal fluid drainage procedures. Part Number: 826616
- Product code
- OJG Neurological Tray
- Reason for recall
- Incorrect extended expiration dates identified on the outer box label only, if product is used past the correct expiration date there is a potential for pain, infection, or adverse tissue reaction to occur
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- Integra issued Urgent: Voluntary Medical Deice Correction letter to Distributors and Hospitals dated 1/21/22 stating reason for recall, health risk and action to take: 1. Please review and understand the information provided in this letter. 2. Please ensure you follow the Instructions for Use and use the product by the expiration dates on the labels within the kit for remaining product on your shelf; these expiration dates are all correct as labeled. 3. Please strike out the incorrect expiration dates on the outer box label of the affected product and write the correct expiration dates per Table 1. 4. Complete the attached Acknowledgement Form (even if you have no product on hand) and return the completed form to [email protected] or FAX to 1-609-750-4220 indicating receipt and review of this notification. Keep a copy of the form for your records. We recommend you also maintain a copy of this notification and the signed copy of the acknowledgement form for your records. Regulatory agencies such as the FDA perform audits of field actions of this nature to verify that our customers have been notified and understand the nature of the field action. Should you have any questions regarding these instructions, please contact Customer Service: Monday to Friday 8:00 AM 8:00 PM EST Phone: 1-800-654-2873 Email: [email protected]
- Quantity in commerce
- 1160 kits
- Distribution
- Worldwide distribution - US Nationwide and the country of Hong Kong.
- Product codes and lots
- Lot Numbers: 20DDA264, 21HDA861 UDI# 10381780520382
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0817-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0818-2022 | Class II | Terminated | Voluntary: Firm initiated |