Z-0821-2022Class II
Compella Therapy Air Supply Unit
Baxter Healthcare Corporation / initiated 2022-02-14 / Terminated / root cause: labeling and UDI
Baxter Healthcare Corporation began a recall of Compella Therapy Air Supply Unit on . FDA classified it as Class II and gives the reason as "Potential damage to the power cord may cause the power conductors to be exposed, thereby creating a risk of shock from the power mains". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0821-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Compella Therapy Air Supply Unit
- Product code
- IOQ Bed, Flotation Therapy, Powered
- Reason for recall
- Potential damage to the power cord may cause the power conductors to be exposed, thereby creating a risk of shock from the power mains.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- Customer notification letters were mailed beginning 02/21/2022. A localized customer letter will be mailed (with signature service) to all first-level consignees by Sedgwick. The letter will inform the customer of the issue, reason for Field Action and provide risk mitigations. The customer will be instructed to sign acknowledgement form to confirm receipt and return to a dedicated email address. Distributors will be asked to forward to their end users. Distributors will be provided with instructions on communication with their end users. All consignees will be required to return a signed acknowledgment form and confirm that they read and understand the issue. Distributors will be asked to forward to their end users. Distributors will be provided instruction on communication with their end users.
- Quantity in commerce
- 6022 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Brazil, Cameroon, Canada, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, India, Ireland, Italy, Jamaica, Jordan, Kuwait, Mexico, Netherlands, New Zealand, Norway, Oman, Poland, Qatar, Romania, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkmenistan, UAE, and UK.
- Product codes and lots
- All lots/serial numbers UDI information: Compella Therapy Air Supply Unit - 00887761013803
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0820-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0821-2022 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.