Z-0830-2022Class II

Siemens Dimension Vista Magnesium Flex reagent cartridge (MG) Material # 10445158, Catalog # K3057 used in...

Siemens Healthcare Diagnostics Inc. / initiated 2022-01-28 / Open, Classified

Siemens Healthcare Diagnostics, Inc. began a recall of Siemens Dimension Vista Magnesium Flex reagent cartridge (MG) Material # 10445158, Catalog # K3057 used in the Dimension Vista System. The Dimension Vista... on . FDA classified it as Class II and gives the reason as "Dimension Vista Magnesium Flex reagent cartridge lots exhibiting imprecision and intermittent Abnormal Assay flags. When this issue occurs, there is a potential for QC failures or erroneous patient results which may...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0830-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens Dimension Vista Magnesium Flex reagent cartridge (MG) Material # 10445158, Catalog # K3057 used in the Dimension Vista System. The Dimension Vista Magnesium Flex Reagent (MG) method is an In vitro diagnostic test for the quantitative measurement of magnesium in human serum, plasma, and urine on the Dimension Vista System. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypomagnesemia (abnormally high plasma levels of magnesium).
Product code
JGJ Photometric Method, Magnesium
Reason for recall
Dimension Vista Magnesium Flex reagent cartridge lots exhibiting imprecision and intermittent Abnormal Assay flags. When this issue occurs, there is a potential for QC failures or erroneous patient results which may lead to repeat and follow-up testing.
FDA root cause
Under Investigation by firm
Action
The firm, Siemens Healthineers, sent "URGENT MEDICAL DEVICE CORRECTION" (VC-22-01.A.US) on 01/28/2022 via Fed Ex to all customers to inform them of the issue and instruct them to discontinue use and discard the affected lots. Complete and return the Field Correction Effectiveness Check Form via email at [email protected] or fax to: (312) 275-7795. At the present time, Siemens has no replacement lots to offer customers. Siemens suggests using an alternate methodology for magnesium until a replacement lot is available. If you received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Remote Services Center or your local Siemens Technical Support Representative. Replacement kits will be provided when a new lot is available. If you have any questions, please contact your Siemens Healthineers Remote Services Center or your local Siemens Healthineers Technical Support Representative.
Quantity in commerce
14220 (US 13305, OUS 915)
Distribution
Worldwide distribution: US (nationwide) 47 States including: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY) and OUS countries of: Australia, Canada and Netherlands.
Product codes and lots
[lot 21253BB, UDI: 0084276801569421253BB22091010445158840], [lot 21263BA UDI: 0084276801569421263BA22092010445158840], [lot 21335AA, UDI: 0084276801569421335AA22120110445158840]

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0830-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0830-2022Class IIOngoingVoluntary: Firm initiated