Z-0870-2022Class II

Evita V500 Ventilator with Software 2.51.01 and Lower - intended for the ventilation of adult, pediatric and...

Draeger, Inc. / initiated 2022-02-14 / Terminated / root cause: manufacturing process

Draeger Medical, Inc. began a recall of Evita V500 Ventilator with Software 2.51.01 and Lower - intended for the ventilation of adult, pediatric and neonatal patients Catalog Number: 8416400 on . FDA classified it as Class II and gives the reason as "Cybersecurity Improvement Action-not equipped against potential cyber security threats". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0870-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Draeger, Inc.
Product
Evita V500 Ventilator with Software 2.51.01 and Lower - intended for the ventilation of adult, pediatric and neonatal patients Catalog Number: 8416400
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Cybersecurity Improvement Action-not equipped against potential cyber security threats
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Draeger issued "Urgent Medical Device Correction Notices" on 2/14/22 to the Director, Clinical Engineering and Risk Manager via FedEx with tracking/confirmation of delivery. Letter states reason for recall, health risk and action to take: Continue to follow the recommendations in the instructions for use: o Limit or control physical access to the abovementioned ventilators. o Do not connect any non-approved devices to the USB, LAN and DVI interfaces. o Be attentive to notifications, alarms, and alerts Consider closing/covering all unused USB, LAN and DVI interfaces Should you decide to close/cover unused interfaces, Draeger offers to provide tools upon request to cover or close these data interfaces of the ventilators free of charge. For the intended and authorized use of the interfaces the USB locks and interface covers can be removed with the provided tool. To order this material please contact Dragger Customer Success via phone between the hours of 8:00 AM 8:00 PM EST at 1-800-437-2437 (press 1 at the prompt, the 2), via fax at 1-215-723-5935, or via email at [email protected]. You will need to know the material number to order, which is dependent on the model of the C500 Cockpit that is installed on your V500/VN500 Ventilator. If you have any questions regarding the operation of your Evita V500 or Babylog VN500, please contact Drager Service Technical Support between the hours of 8:00 AM 8:00 PM EST at 1-800-437-2437 (press 2 at the prompt, then 2, then 1).
Quantity in commerce
24,337 (10,728 units US)
Distribution
US Nationwide
Product codes and lots
All Lots/Serial Numbers. UDI: 04048675042266

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0870-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0870-2022Class IITerminatedVoluntary: Firm initiated
Z-0871-2022Class IITerminatedVoluntary: Firm initiated