Z-0859-2022Class II

SIGNA Excite 3T. Magnetic Resonance Imaging System

GE Healthcare / initiated 2022-02-18 / Open, Classified / root cause: design

GE Healthcare, LLC began a recall of SIGNA Excite 3T. Magnetic Resonance Imaging System on . FDA classified it as Class II and gives the reason as "There is potential for the images to be flipped left to right". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0859-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Healthcare
Product
SIGNA Excite 3T. Magnetic Resonance Imaging System
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
There is potential for the images to be flipped left to right.
Root cause tag
design
FDA root cause
Device Design
Action
This issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected systems. This 2/18/2022 letter informs the customer of the issue and provides actions to be taken for the continued use of the device. GE Healthcare will provide a revised service manual that includes a required Finalization step that ensures a geometry check with a DQA phantom (or similar) following Gradient Switch servicing to ensure cable placement is correct and will have no impact on image orientation. A second letter was send to customers on 07/27/2022 and included instructions and link for customer to access and download the revised service manual.
Quantity in commerce
55 units
Distribution
Worldwide Distribution. US nationwide, Argentina, Australia, Azerbaijan, Bosnia and Herzegovina, Brazil, Canada, China, Denmark, Ecuador, Ethiopia, France, Germany, Greece, Hong Kong, India, Iran, Ireland, Italy, Japan, Korea, Kuwait, Lebanon, Mexico, Netherlands, New Zealand, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Taiwan, Turkey, United Arab Emirates, and United Kingdom.
Product codes and lots
System IDs: 3422, 209577MRMOBX, 407303CHMR, SIMONVILLAGES3T, 321MR3T3, 321MR3T2, 407999MR3T, 316268TWIN3, 8109873T, 732473MR, 702433SIM3T, 864573MBMR2, 901726OPMR, 214879INVIVO, 817578LGMR2, 972665XDDAMR1 2284SIG15T, 47413, 2550000, 975, 604270MR3T, 416813MR3, 416369MR2, 082427130025 082427130020, 082427130022, 082427270012, 082427270011, 082427150025, 082427020020, 082427010032, 082427160051, 082427160050, 082427120042, 082427120027, 082427030037, 082427250016, 082427070046, 082427150026, 082427090035, 082427140034, 082427140040, 082427090044, 082427100075, 082427120029, 082427070049, 082427070048, 082427220010, 082427300008, YM1408, YM1414, YM1406, 850270152, LB1049MR02, PK1184MR01

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0859-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0857-2022Class IIOngoingVoluntary: Firm initiated
Z-0858-2022Class IIOngoingVoluntary: Firm initiated
Z-0859-2022Class IIOngoingVoluntary: Firm initiated
Z-0860-2022Class IIOngoingVoluntary: Firm initiated