Z-0894-2022Class II

BenchMark ULTRA and DISCOVERY ULTRA Instruments

Ventana Medical Systems Inc / initiated 2022-01-20 / Open, Classified / root cause: design

Ventana Medical Systems Inc began a recall of BenchMark ULTRA and DISCOVERY ULTRA Instruments on . FDA classified it as Class II and gives the reason as "Potential for Fluid Leak inside a staining system that could cause an electrical short circuit of the General Purpose Input/ Output Board's J1 Connector that could result in smoke and fire". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0894-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Ventana Medical Systems Inc
Product
BenchMark ULTRA and DISCOVERY ULTRA Instruments
Product code
KPA Slide Stainer, Automated
Reason for recall
Potential for Fluid Leak inside a staining system that could cause an electrical short circuit of the General Purpose Input/ Output Board's J1 Connector that could result in smoke and fire.
Root cause tag
design
FDA root cause
Device Design
Action
On 01/20/22, correction notices were sent to customers who were asked to take the following actions needed to continue affected device use: 1) Inform any operators of the potential hazards associated with this issue and provide a copy of this notification as appropriate. 2) Inspect the instrument daily for evidence of leaks (i.e., liquid on the floor, or in and around the instrument). 3) If you experience a fluid leak, stop using the system and turn off the power to the instrument. Immediately contact the firm's Network Customer Support Center at 1-800-227-2155. 4) Complete the attached fax back form and return via fax or email. 5) If your facility has distributed the affected product to another site, please ensure this UMDC is provided to that site. 6) File this UMDC for future reference. The firm will send a representative to schedule a visit to perform an inspection, reroute cables, and replace parts as necessary to help mitigate the issue. Customers with questions were encouraged to call the Customer Support Center at the before mentioned number; support is available 24 hours a day, seven days a week. On 4/19/23, correction notices were distributed to additional customers who may have received devices without a drip loop, which is the solution to this event. Customers were told they would be contacted by a Service Representative to coordinate specific service actions, including the inspection of GPIO boards and rerouting cables - drip loop introduction.
Quantity in commerce
9214
Distribution
US: CA, IL, PA, FL, AZ, NY, OH, CO, AR, NC, MD, AL, AK, LA, IA, MN, MA, ND, IN, TX, TN, CT, SC, GA, UT, WI, VA, SD, KY, MS, MO, DE, MI, MT, NJ, WV, OK, WA, OR, PR, DC, NM, ID, NH, NE, ME, KS, HI, NV, RI, WY, VT OUS: Australia, Austria, Belgium, Canada, China, Croatia, Denmark, Finland, France, Germany, Hong Kong, Hungary, India, Israel, Italy, Japan, Netherlands, Norway, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, United Kingdom, Algeria, Argentina, Aruba, Bosnia-Herzegovina., Brazil, Bulgaria, Chile, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Greece, Indonesia, Ireland, Kazakhstan, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Macedonia, Malaysia, Martinique, Mexico, Monaco, Morocco, Namibia, New Zealand, Oman, Pakistan, Panama, Peru, Philippines, Reunion, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, Vietnam, White Russia, Zimbabwe
Product codes and lots
Serial Numbers Less-Than-Or-Equal-To 322030

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0894-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0894-2022Class IIOngoingVoluntary: Firm initiated