Z-0747-2022Class II

ProteusPLUS and ProteusONE- designed to produce and deliver a proton beam for the treatment of patients with...

Ion Beam Applications S.A. / initiated 2022-02-15 / Open, Classified / root cause: design

Ion Beam Applications S.A. began a recall of ProteusPLUS and ProteusONE- designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible... on . FDA classified it as Class II and gives the reason as "When resuming from a beam pause, the Proton Therapy System does not verify whether the beam range has not been manually modified during the pause and is still the prescribed one. Therefore, if an operator manually...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0747-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Ion Beam Applications S.A.
Product
ProteusPLUS and ProteusONE- designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation Proteus 235
Product code
LHN System, Radiation Therapy, Charged-Particle, Medical
Reason for recall
When resuming from a beam pause, the Proton Therapy System does not verify whether the beam range has not been manually modified during the pause and is still the prescribed one. Therefore, if an operator manually modified the range during a pause, there is a risk that a portion of the treatment beam after the resume is delivered with an error in range
Root cause tag
design
FDA root cause
Device Design
Action
IBA issued Urgent Medical Device Correction notification on February 15, 2022 directly to customer either by hand or by email. Letter states reason for recall, health risk and action to take: For Proton Therapy sites where the system is operated by the customer: - Warn operators to not manually change the beam range (action called manual set range ) while a patient is on the Patient Positioning System. - IBA will provide a training to all operators about the correct use of the action called manual set range and the risk associated to its use while a patient is on the Patient Positioning System. This training will be provided on your site by August 31, 2022. Contact: [email protected]. Help Desk: +32 2 507 20 81 (available 24/7)
Quantity in commerce
15 units US; 22 units OUS
Distribution
IL, KS, LA, FL, MI, OK, PA, NJ, TN, TX, VA, WA Foreign: Belgium, England, Germany, France, Czech Republic, Italy, Poland, Russia, Spain, Sweden, South Korea, Taiwan, Netherlands,
Product codes and lots
PAT.003 (KR), PAT.006 (US), PAT.107 (EU), PAT.108 (US), PAT109 (US), PAT.110 (US), PAT.111 (EU), PAT.112 (US), PAT.113 (US), PAT.114 (EU), PAT.115 (EU), SAT.116 (US), SAT.117 (EU), SAT.118 (RU), SAT.119 (US), SAT.120 (EU), SAT.122 (EU), SAT.123 (US), SAT.125 (IN), SAT.126 (US), SAT.127 (TW), SAT.132 (EU), SAT.133 (US), SAT.136 (IN), SAT.140 (US), SBF.101 (EU), SBF.103 (JP), SBF.104 (JP), SBF.105 (US), SBF.107 (FR), SBF.109 (UK), SBF.110 (UK), SBF.112 (EU), SBF.113 (US), SBF.115 (UK), SBF.117 (EU), SBF.128 (US).  UDI: (01)05404013801138

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0747-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0747-2022Class IIOngoingVoluntary: Firm initiated