Z-0730-2022Class II
YSIO X.Pree
Siemens Medical Solutions USA, Inc. / initiated 2022-03-07 / Open, Classified / root cause: software
Siemens Medical Solutions USA, Inc began a recall of YSIO X.Pree on . FDA classified it as Class II and gives the reason as "For the automated multi-image-acquisition procedure Ortho x-ray collimation is set in a preparative stage for the entire examination area prior to the exam. During acquisition of each individual x-ray image, the x-ray...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0730-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- YSIO X.Pree
- Product code
- KPR System, X-Ray, Stationary
- Reason for recall
- For the automated multi-image-acquisition procedure Ortho x-ray collimation is set in a preparative stage for the entire examination area prior to the exam. During acquisition of each individual x-ray image, the x-ray collimator is automatically positioned in a way that the subsequent series of acquisition covers the defined field of view needed for each step. However, during the acquisition the collimation area displayed to the operator on the User Interface does not correctly represent the collimation area specified by the system. It indicates to the user an open collimator instead, e.g., the information displayed on the User Interface shows wider area of collimation than values preset prior to the examination. However, the collimation of the x-ray is performed correctly and always matches the examination area predefined by the user.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice to all affected customers. The letter inform customers that the firm will distribute a software correction via update instruction at no cost to the customers.
- Quantity in commerce
- 53 units
- Distribution
- US Nationwide, Worldwide distribution
- Product codes and lots
- 11107464