Z-0730-2022Class II

YSIO X.Pree

Siemens Medical Solutions USA, Inc. / initiated 2022-03-07 / Open, Classified / root cause: software

Siemens Medical Solutions USA, Inc began a recall of YSIO X.Pree on . FDA classified it as Class II and gives the reason as "For the automated multi-image-acquisition procedure Ortho x-ray collimation is set in a preparative stage for the entire examination area prior to the exam. During acquisition of each individual x-ray image, the x-ray...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0730-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
YSIO X.Pree
Product code
KPR System, X-Ray, Stationary
Reason for recall
For the automated multi-image-acquisition procedure Ortho x-ray collimation is set in a preparative stage for the entire examination area prior to the exam. During acquisition of each individual x-ray image, the x-ray collimator is automatically positioned in a way that the subsequent series of acquisition covers the defined field of view needed for each step. However, during the acquisition the collimation area displayed to the operator on the User Interface does not correctly represent the collimation area specified by the system. It indicates to the user an open collimator instead, e.g., the information displayed on the User Interface shows wider area of collimation than values preset prior to the examination. However, the collimation of the x-ray is performed correctly and always matches the examination area predefined by the user.
Root cause tag
software
FDA root cause
Software design
Action
Siemens Medical Solutions USA, Inc. has initiated a Customer Advisory Notice to all affected customers. The letter inform customers that the firm will distribute a software correction via update instruction at no cost to the customers.
Quantity in commerce
53 units
Distribution
US Nationwide, Worldwide distribution
Product codes and lots
11107464

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0730-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0730-2022Class IIOngoingFDA Mandated