Z-0734-2022Class II

Azurion systems with software release R2.0.x

Philips North America LLC / initiated 2021-05-26 / Completed / root cause: software

Philips Healthcare began a recall of Azurion systems with software release R2.0.x on . FDA classified it as Class II and gives the reason as "In the Azurion system, the user can add a new study to a patient by selecting the option "Add Study". The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a...". As of , the recall status on record is Completed, and MDM tags the root cause as software.

Recall record

Recall number
Z-0734-2022
Classification
Class II
Status
Completed
Initiated
Posted
Recalling firm
Philips North America LLC
Product
Azurion systems with software release R2.0.x
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
In the Azurion system, the user can add a new study to a patient by selecting the option "Add Study". The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a software defect, when the study is initiated by pressing Start Procedure , the Patient Type changes inadvertently to a Patient Type different than the one selected as shown in the Table below. Patient type is one of the factors involved in the dose control process. The incorrect patient type changes the technique factors to be used by the system without notification to the user.
Root cause tag
software
FDA root cause
Radiation Control for Health and Safety Act
Action
Philips notified customers of the issue via an Electronic Product Radiation Defect notification letter dated February 8, 2022. Philips Field Service Engineers will schedule an appointment with customers to install the software update. Philips will correct the affected systems at no cost to customers with a software upgrade. For information, contact your local Philips representative: 1-800-722-9377.
Quantity in commerce
69 systems
Distribution
US Nationwide Distribution
Product codes and lots
Model(s) Azurion R2.1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0734-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0734-2022Class IICompletedFDA Mandated