Z-0737-2022Class II

Philips StentBoost Live R2.0 application, when used with the Philips Azurion with software release R2.x

Philips North America LLC / initiated 2022-02-02 / Terminated

Philips Healthcare began a recall of Philips StentBoost Live R2.0 application, when used with the Philips Azurion with software release R2.x on . FDA classified it as Class II and gives the reason as "Due to an incorrect configuration setting in the EPX database of StentBoost Live, acquisition does not stop automatically as it should. When the user selects the StentBoost Live protocol on the Philips Azurion system...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0737-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips StentBoost Live R2.0 application, when used with the Philips Azurion with software release R2.x
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Due to an incorrect configuration setting in the EPX database of StentBoost Live, acquisition does not stop automatically as it should. When the user selects the StentBoost Live protocol on the Philips Azurion system, the following on-screen message is displayed to the user in the Philips StentBoost Live application: "Press the cine pedal until the acquisition stops". However, due to the incorrect configuration of the EPX database, the acquisition does not stop after 40 images, but instead it continues as long as the pedal is pressed.
FDA root cause
Radiation Control for Health and Safety Act
Action
Philips notified customers of the issue via an Electronic Product Radiation Defect notification letter dated February 2, 2022. Philips will schedule an appointment with customers to install the software update. Philips will correct the affected systems at no cost to customers with a software upgrade. For information contact your local Philips representative: 1-800-722-9377.
Quantity in commerce
42 Systems
Distribution
US Nationwide distribution.
Product codes and lots
Azurion R2.1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0737-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0737-2022Class IITerminatedFDA Mandated