Z-0815-2022Class II
Revolution Apex, Revolution CT with Apex Edition
GE Healthcare / initiated 2021-06-30 / Open, Classified
GE Healthcare, LLC began a recall of Revolution Apex, Revolution CT with Apex Edition on . FDA classified it as Class II and gives the reason as "The accumulated dose is incorrectly displayed in certain situations. There are two situations that will result in the SmartStep displayed dose being inaccurately high". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0815-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- GE Healthcare
- Product
- Revolution Apex, Revolution CT with Apex Edition
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- The accumulated dose is incorrectly displayed in certain situations. There are two situations that will result in the SmartStep displayed dose being inaccurately high.
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- GE Healthcare sent an Important Electronic Product Radiation Warning letter dated February 8, 2022 to all affected systems informing its users of the potential for the issue to occur and instruct the users as to what actions to take. GE Healthcare will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services, the details of which will be included in a subsequent communication to you or through a GE Healthcare field engineer site visit. For questions contact GE Healthcare Services at 1 (800) 437-1171 or local Service Representative.
- Quantity in commerce
- 273 (51 US, 222 OUS) in total
- Distribution
- Worldwide - US Nationwide distribution.
- Product codes and lots
- Revolution Apex, Revolution CT with Apex Edition
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0815-2022 | Class II | Ongoing | FDA Mandated | ||
| Z-0816-2022 | Class II | Ongoing | FDA Mandated |