Z-0816-2022Class II

Revolution CT, Revolution CT ES

GE Healthcare / initiated 2021-06-30 / Open, Classified

GE Healthcare, LLC began a recall of Revolution CT, Revolution CT ES on . FDA classified it as Class II and gives the reason as "The accumulated dose is incorrectly displayed in certain situations. There are two situations that will result in the SmartStep displayed dose being inaccurately high". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0816-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Healthcare
Product
Revolution CT, Revolution CT ES
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
The accumulated dose is incorrectly displayed in certain situations. There are two situations that will result in the SmartStep displayed dose being inaccurately high.
FDA root cause
Radiation Control for Health and Safety Act
Action
GE Healthcare sent an Important Electronic Product Radiation Warning letter dated February 8, 2022 to all affected systems informing its users of the potential for the issue to occur and instruct the users as to what actions to take. GE Healthcare will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services, the details of which will be included in a subsequent communication to you or through a GE Healthcare field engineer site visit. For questions contact GE Healthcare Services at 1 (800) 437-1171 or local Service Representative.
Quantity in commerce
273 (51 US; 222 OUS) in total
Distribution
Worldwide - US Nationwide distribution.
Product codes and lots
Revolution CT, Revolution CT ES

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0816-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0815-2022Class IIOngoingFDA Mandated
Z-0816-2022Class IIOngoingFDA Mandated