Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0943-2022Class II

Breas Ventilator Trolley with the Vivo 50/60/65 and Vivo 45LS ventilator models Article no 007384

Breas Medical AB / initiated 2022-03-21 / Terminated / root cause: manufacturing process

Breas Medical, Inc. began a recall of Breas Ventilator Trolley with the Vivo 50/60/65 and Vivo 45LS ventilator models Article no 007384 on . FDA classified it as Class II and gives the reason as "Ventilator mounting brackets for the Vivo 50/60/65 and Vivo 45LS ventilator models can become detached from the Breas Ventilator Trolley, this may cause the ventilator to fall from the trolley and cause injury to a...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0943-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Breas Medical AB
Product
Breas Ventilator Trolley with the Vivo 50/60/65 and Vivo 45LS ventilator models Article no 007384
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Ventilator mounting brackets for the Vivo 50/60/65 and Vivo 45LS ventilator models can become detached from the Breas Ventilator Trolley, this may cause the ventilator to fall from the trolley and cause injury to a person who is in the immediate proximity
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Breas Medical issued recall information via telephone followed by Urgent Medical Device Recall letter to Distributors/End Users via email on 3/21/22 via telephone.. Letter state reason for recall, health risk and action to take: Breas Medical s solution is to provide Inspect and Repair Kits (REF 008591) to all those potentially affected by this notice. The Inspect and Repair Kits are used to inspect the Trolley to determine if remedial action is required, and to replace the defective part if needed. We will send you enough Inspect and Repair Kits (REF 008591)s for all trolleys in your possession. " We (the Customer) will identify Breas Ventilator Trolleys (REF 007384) delivered between June 11th, 2020, and December 3rd, 2021. " Breas will provide us with the corresponding number of Inspect and Repair Kits (REF 008591). " We will inspect the swivel assembly and complete the enclosed form to indicate whether the trolley contained the necessary washer and no further action is needed, or if the washer was missing and the swivel assembly was replaced by the new one. " We will scrap the replaced swivel assembly. e kindly ask you to acknowledge receipt of this Field Safety Notice together with your choice to remedy the issue using attached reply form Trolley Field Safety Notice Distributor/Customer Reply Form with Option Selection . Contact Information Email [email protected] Helpline +1 (855) 436-8724 Postal Address 16 Esquire Road, N. Billerica, MA 01862 Web Portal breas.com
Quantity in commerce
707 units
Distribution
US Nationwide
Product codes and lots
Trolleys delivered between June 11th, 2020, and December 3rd, 2021. UDI: 07321820073849

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0943-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0943-2022Class IITerminatedVoluntary: Firm initiated