Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (END) - VERSION...
B Braun Medical Inc began a recall of Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (END) - VERSION 2, Article Code Number 3456103A. These... on . FDA classified it as Class II and gives the reason as "Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0873-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- B Braun Medical Inc
- Product
- Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (END) - VERSION 2, Article Code Number 3456103A. These conductivity sensors may be used as spare parts for the following dialysis machines: 710200L (UDI 04046963686345), 710200S (UDI 04046964285608), 710200U (UDI 04046964285615), 710200K (UDI 04046963686338), 710500L (UDI Not applicable) and 710500K (UDI Not applicable).
- Product code
- KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
- Reason for recall
- Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration depending on whether the dialysis machine is equipped with a Dialysis Fluid filter.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On February 17, 2022, the firm, B.Braun Medical Inc., mailed out Recall Customer Notification Letters to affected consignees. Distributors should notify affected customers. Customers should identify any affected products within the spare parts inventory of their facility, cease use, and quarantine the product. Do not destroy any affected product. If any machine exhibited repeated Alarm Code 1026 - "UF balance? Air leakage in dialyzer coupling" alarms, customer should discontinue therapy on the device and immediately quarantine the machine until it can be inspected and any impacted sensors replaced. Customers should inspect ALL Dialog+ and Dialog+ Evolution dialysis machines in their facility which may have been serviced with potentially impacted sensors. Once the firm receives the customer acknowledgement form, a representative will contact the customer on how to return impacted sensors and/or to schedule a time to inspect equipment. Customers with questions should contact B. Braun Avitum Technical Services at 1-800-21-0445.
- Quantity in commerce
- 95
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Lots 4/21 and 5/21
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0872-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0873-2022 | Class II | Terminated | Voluntary: Firm initiated |