Gallant DR Implantable Cardioverter Defibrillator REF # CDDRA500Q
St. Jude Medical, Cardiac Rhythm Management Division began a recall of Gallant DR Implantable Cardioverter Defibrillator REF # CDDRA500Q on . FDA classified it as Class II and gives the reason as "Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0988-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abbott Medical
- Product
- Gallant DR Implantable Cardioverter Defibrillator REF # CDDRA500Q
- Product code
- LWS Implantable Cardioverter Defibrillator (Non-Crt)
- Reason for recall
- Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Abbott issued a Safety Notification For a subset of Gallant ICDS and CRT-D, Models CDVRA500Q, CDDRA500Q and CDHFA500Q on March 10, 2022. Notification was hand delivered in US. The notification includes the Patient Management Recommendations: -As the devices are unable to be followed using remote monitoring via Bluetooth connectivity, follow-ups should be conducted in-clinic using inductive telemetry. - Abbott has developed a process to attempt recovery of Bluetooth communication via a device firmware upgrade. If remote monitoring via Bluetooth communication is desired, please work with your local Abbott representative to schedule a firmware upgrade. Please complete and return the included Acknowledgement Form to [email protected] and maintain a record of this notice to ensure effectiveness of the communication. For any question, please contact Abbott Technical Support at 1-800-722-3774 (U.S.).
- Quantity in commerce
- 1
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of FL, KY and the country of France, Czech Republic.
- Product codes and lots
- REF# CDDRA500Q SN# 111018237, Exp Date: 31-Oct-2022 Material# 600115496 UDI# 05415067032003
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0988-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0989-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0990-2022 | Class II | Ongoing | Voluntary: Firm initiated |