Z-0990-2022Class II

Gallant HF Implantable Cardioverter Defibrillator REF# CDHFA500Q

Abbott Medical / initiated 2022-03-10 / Open, Classified / root cause: design

St. Jude Medical, Cardiac Rhythm Management Division began a recall of Gallant HF Implantable Cardioverter Defibrillator REF# CDHFA500Q on . FDA classified it as Class II and gives the reason as "Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0990-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Abbott Medical
Product
Gallant HF Implantable Cardioverter Defibrillator REF# CDHFA500Q
Product code
LWS Implantable Cardioverter Defibrillator (Non-Crt)
Reason for recall
Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.
Root cause tag
design
FDA root cause
Component design/selection
Action
Abbott issued a Safety Notification For a subset of Gallant ICDS and CRT-D, Models CDVRA500Q, CDDRA500Q and CDHFA500Q on March 10, 2022. Notification was hand delivered in US. The notification includes the Patient Management Recommendations: -As the devices are unable to be followed using remote monitoring via Bluetooth connectivity, follow-ups should be conducted in-clinic using inductive telemetry. - Abbott has developed a process to attempt recovery of Bluetooth communication via a device firmware upgrade. If remote monitoring via Bluetooth communication is desired, please work with your local Abbott representative to schedule a firmware upgrade. Please complete and return the included Acknowledgement Form to [email protected] and maintain a record of this notice to ensure effectiveness of the communication. For any question, please contact Abbott Technical Support at 1-800-722-3774 (U.S.).
Quantity in commerce
1
Distribution
Worldwide distribution - US Nationwide distribution in the states of FL, KY and the country of France, Czech Republic.
Product codes and lots
REF# CDHFA500Q SN #111004346, Exp Date : 30-Jun-2022 Material # 100173746 UDI: 05415067032010

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0990-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0988-2022Class IIOngoingVoluntary: Firm initiated
Z-0989-2022Class IIOngoingVoluntary: Firm initiated
Z-0990-2022Class IIOngoingVoluntary: Firm initiated