Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0972-2022Class II

Artis zee biplane MN, Model Number 10094143

Siemens Medical Solutions USA, Inc. / initiated 2022-02-23 / Open, Classified / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Artis zee biplane MN, Model Number 10094143 on . FDA classified it as Class II and gives the reason as "If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0972-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Artis zee biplane MN, Model Number 10094143
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".
Root cause tag
design
FDA root cause
Device Design
Action
An URGENT: MEDICAL DEVICE CORRECTION notification letter dated 2/23/22 was sent to customers. Which steps have to be taken by the user to avoid the possible risks associated with this problem? As also described in the Operator Manual, we recommend that the system operator checks the water level of the cooling circuit at least every three months and refills water, if necessary: 1. Open the filling gland of the cooling unit. The water surface must be clearly visible above the cooling ribs. 2. Replenish with water (drinking water quality) if cooling liquid is lacking. Please inform the service technician in case of lacking cooling liquid. How will the corrective action be implemented and what is the efficiency of the corrective actions? The tube cooling unit will be equipped with a water level sensor and the software will be updated accordingly. In case the coolant level drops below a certain threshold the following message will be displayed: Tube cooling water level low: Refill as per Operator Manual . Siemens will correct the error with a hardware and software update via Update Instruction AX002/20/S and AX004/20/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. We do not consider it necessary to re-examine any patients in relation to the issues described above. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented. Acknowledge Receipt of this Safety Advisory Notice Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein. If you have received this notification by email via Adobe EchoSign, kindly sign with your digital signature to acknowledge that you have read and understand t
Quantity in commerce
7 systems
Distribution
US Nationwide.
Product codes and lots
UDI: 4056869010106; Serial Numbers: 149100 149401 149421 149106 149202 149420 149402

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0972-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0967-2022Class IIOngoingVoluntary: Firm initiated
Z-0968-2022Class IIOngoingVoluntary: Firm initiated
Z-0969-2022Class IIOngoingVoluntary: Firm initiated
Z-0970-2022Class IIOngoingVoluntary: Firm initiated
Z-0971-2022Class IIOngoingVoluntary: Firm initiated
Z-0972-2022Class IIOngoingVoluntary: Firm initiated
Z-0973-2022Class IIOngoingVoluntary: Firm initiated
Z-0974-2022Class IIOngoingVoluntary: Firm initiated
Z-0975-2022Class IIOngoingVoluntary: Firm initiated
Z-0976-2022Class IIOngoingVoluntary: Firm initiated
Z-0977-2022Class IIOngoingVoluntary: Firm initiated
Z-0978-2022Class IIOngoingVoluntary: Firm initiated
Z-0979-2022Class IIOngoingVoluntary: Firm initiated
Z-0980-2022Class IIOngoingVoluntary: Firm initiated