Z-0987-2022Class II

Artis pheno- in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended...

Siemens Medical Solutions USA, Inc. / initiated 2022-02-25 / Open, Classified / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Artis pheno- in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use-single and biplane diagnostic imaging and... on . FDA classified it as Class II and gives the reason as "Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). The potential software errors are related to the block...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0987-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Artis pheno- in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use-single and biplane diagnostic imaging and interventional procedures. Model: 10849000
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). The potential software errors are related to the block movements function and manual rotation of the flat detector. Depending on the status of the intervention, the limited functionality may not be sufficient to continue with treatment as planned. This may result in a situation where clinical treatment may be delayed until an additional restart is performed
Root cause tag
software
FDA root cause
Software Design Change
Action
Siemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice to all affected customers via AX005/22/S on 2/28/22. Siemens will provide a software update for all affected systems via Update Instruction AX003/22/S and AX004/22/S. Letter states reason for recall, health risk and action to take: We do not consider it necessary to re-examine any patients in relation to the issues (issue 1 to 5) described above. If measurements have already been performed in the past for diagnostics, please verify the results and diagnostic evaluation if applicable. Siemens will correct the software error via Update Instruction AX003/22/S and Update Instruction AX004/22/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented. Acknowledge Receipt of this Safety Advisory Notice Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein. If you have received this notification by email via Adobe EchoSign, kindly sign with your digital signature to acknowledge that you have read and understand the content provided. If you have received this notification via Certified Mail, please provide at least two contact names and email addresses of the appropriate personnel within your facility who should receive these notifications electronically. This information will assist in providing you with these important notifications in a more expedient and compliant manner. Please include your Siemens system names, material numbers, serial numbers, and Siemens Functional Location numbers and email this information to: recallsandrefusals.team@siemens-hea
Quantity in commerce
20 units
Distribution
US Nationwide distribution.
Product codes and lots
Serial Number:  164719 164712 164740 164722 164763 164735 164723 164743 164741 164728 164711 164718 164739 164703 164721 164726 164724 164725 164760  UDI: 04056869046877

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0987-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0985-2022Class IIOngoingVoluntary: Firm initiated
Z-0986-2022Class IIOngoingVoluntary: Firm initiated
Z-0987-2022Class IIOngoingVoluntary: Firm initiated