Z-0956-2022Class II
Abbott SE TactiCath Sensor Enabled Contact Force Ablation Catheter, Bi-D, Curve D-F, Model A-TCSE-DF, Sterile
Abbott Medical / initiated 2022-03-24 / Open, Classified / root cause: manufacturing process
Abbott began a recall of Abbott SE TactiCath Sensor Enabled Contact Force Ablation Catheter, Bi-D, Curve D-F, Model A-TCSE-DF, Sterile on . FDA classified it as Class II and gives the reason as "Devices were manufactured with a "First Use Date" preprogrammed in the EEPROM which disables the use of the single-use device". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0956-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abbott Medical
- Product
- Abbott SE TactiCath Sensor Enabled Contact Force Ablation Catheter, Bi-D, Curve D-F, Model A-TCSE-DF, Sterile.
- Product code
- OAE Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Fibrillation
- Reason for recall
- Devices were manufactured with a "First Use Date" preprogrammed in the EEPROM which disables the use of the single-use device.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The recalling firm issued letters dated 3/24/2022 on 3/24/22 via hand-delivery. The letters explained the issue, the impact and associated risks, and the steps to reduce risk. The steps included not using any of the remaining inventory of the affected lots listed in Appendix A, complete and return the accompanying Acknowledgment Form, and return all remaining unused affected devices to the recalling firm. The consignee is informed their Abbott representative can assist in returning these devices and obtaining replacements. The notice is to be forwarded to anyone within their organization who may need to be notified and a copy of the notice should be kept along with the completed Acknowledgment Form for their records.
- Quantity in commerce
- 2,134 devices
- Distribution
- Worldwide distribution - US Nationwide - there was government distribution but no military. The countries of Austria, Belgium, Canada, Denmark, France, Germany, Greece, Israel, Italy, Japan, Netherlands, Norway, Palestine, Poland, Portugal, Russia, Slovakia, Spain, Sweden, Taiwan, and United Kingdom.
- Product codes and lots
- Lot numbers: 8162700, 8164687, 8164730, 8172360, 8173365, 8173840, 8175413, 8177068, 8177349, 8177442, 8179358, 8179499, 8180723, 8180900, 8181043, 8182658, 8183581, 8183595, 8184393, 8184739, 8184817, 8186470, 8186747, 8186860, 8188124, 8188351, 8188844, 8190306, 8190438, 8191524, 8192181, 8192250, 8194325, 8195968, 8196343, 8196367, and 8197913. UDI (01)05415067027641(17)230331(10)xxxxxxx (lot number).
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0956-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0957-2022 | Class II | Terminated | Voluntary: Firm initiated |