Z-1037-2022Class II
1) ACS Arthroscopy Kit 2) ACS Knee Kit 3) ACS Hand Pack
American Contract Systems, Inc. / initiated 2022-03-22 / Open, Classified / root cause: manufacturing process
American Contract Systems, Inc. began a recall of 1) ACS Arthroscopy Kit 2) ACS Knee Kit 3) ACS Hand Pack on . FDA classified it as Class II and gives the reason as "Product was sterilized with a higher than specification EO concentration". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1037-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- American Contract Systems, Inc.
- Product
- 1) ACS Arthroscopy Kit 2) ACS Knee Kit 3) ACS Hand Pack
- Product code
- OJH Orthopedic Tray
- Reason for recall
- Product was sterilized with a higher than specification EO concentration.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- A customer letter dated 04/05/2022 was issued to impacted consignees. The letter instructs the following: 1) Review your inventory and segregate affected product. This product should not be shipped out to customers. 2) Return affected product to 2610 NE Industrial Drive, Suite 220, Kansas City, MO 64120. 3) If product was already further distributed, ensure that all downstream customers are properly notified of the recall and maintain records of effectiveness. A Field Action Response Form should be completed and returned to [email protected], even if you do not have product inventory.
- Quantity in commerce
- 114 packs
- Distribution
- US Nationwide distribution in the states of IL, MO, NE, and TX.
- Product codes and lots
- 1) ACS Arthroscopy Kit - Model LMAR15R, Lot 981221; UDI: 00191072152978 2) ACS Knee Kit - Model LWKN46N, Lot 966221; UDI: 00191072150158 3) ACS Hand Pack - Model FHHP63T, Lot 966221, UDI: 00191072149954
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1036-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1037-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1038-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1039-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1040-2022 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY