Z-1015-2022Class II

IMMULITE 2000 Systems HCG [200 Tests, material number 10381206, catalog number L2KCG] and [600 Tests...

Siemens Healthcare Diagnostics Inc. / initiated 2022-02-22 / Open, Classified

Siemens Healthcare Diagnostics, Inc. began a recall of IMMULITE 2000 Systems HCG [200 Tests, material number 10381206, catalog number L2KCG] and [600 Tests, material number 10381194, catalog number L2KCG6] on . FDA classified it as Class II and gives the reason as "Potential for falsely elevated hCG results due to sample carryover in serum and urine patient samples, as well as quality control samples and adjustors". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1015-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
IMMULITE 2000 Systems HCG [200 Tests, material number 10381206, catalog number L2KCG] and [600 Tests, material number 10381194, catalog number L2KCG6]
Product code
DHA System, Test, Human Chorionic Gonadotropin
Reason for recall
Potential for falsely elevated hCG results due to sample carryover in serum and urine patient samples, as well as quality control samples and adjustors.
FDA root cause
Under Investigation by firm
Action
Urgent Medical Device Correction was issued February 02, 2022 via Fed Ex. Notification provides instructions to customers to mitigate this sample carryover. Customers are requested to complete and return the Effectiveness Check questionnaire. Contact your local Siemens Healthineers technical support representative with any questions.
Quantity in commerce
10,755 boxes (533 US, 10,222 OUS)
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, ID, IN, MN, NC, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TX, VA, WI, WY and OUS: 81 countries.
Product codes and lots
8 production lots: 463, 464, 466, 467, 469, 470, 471, 472, [UDI for 200 tests: (01)00630414961132(10)463(17)20220228, (01)00630414961132(10)464(17)20220331, (01)00630414961132(10)466(17)20220331, (01)00630414961132(10)467(17)20220531, (01)00630414961132(10)469(17)20220731, (01)00630414961132(10)470(17)20220930, (01)00630414961132(10)471(17)20221130, (01)00630414961132(10)472(17)20230228] [UDI for 600 tests: (01)00630414961149(10)463(17)20220228, 01)00630414961149(10)464(17)20220331, (01)00630414961149(10)466(17)20220331, (01)00630414961149(10)467(17)20220531, (01)00630414961149(10)469(17)20220731, (01)00630414961149(10)470(17)20220930, (01)00630414961149(10)471(17)20221130, (01)00630414961149(10)472(17)20230228]