Z-0995-2022Class II

syngo.via RT Image Suite upgraded to VB60 running on SOMATOM go.Up Model Number 11061620. Running on servers...

Siemens Medical Solutions USA, Inc. / initiated 2022-03-07 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of syngo.via RT Image Suite upgraded to VB60 running on SOMATOM go.Up Model Number 11061620. Running on servers with syngo.via VB60 software after upgrade and CT... on . FDA classified it as Class II and gives the reason as "Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Marking step". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0995-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
syngo.via RT Image Suite upgraded to VB60 running on SOMATOM go.Up Model Number 11061620. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgrade.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Marking step.
Root cause tag
software
FDA root cause
Software change control
Action
4/4/22 update: Urgent Medical Device Correction letter has not been issued. The proposed letter informs customers of software issue and workaround to resolve issue. It also requires electronic signature to acknowledge receipt of the notification. If you have additional questions contact the service organization at 1-800-888-7436.
Quantity in commerce
1 (US)
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, IL, IN, KY, MD, MO, ND, NH, NY, OH, OR, PA, SD, TX, VA and the countries of Argentina, Austria, Bahrain, Belgium, Brazil, Czechia, Denmark, France, Germany, Hungary, India, Italy, Japan, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Switzerland, United Kingdom.
Product codes and lots
Partial UDI: 04056869051314, Serial Number 112261

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0995-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0994-2022Class IITerminatedVoluntary: Firm initiated
Z-0995-2022Class IITerminatedVoluntary: Firm initiated
Z-0996-2022Class IITerminatedVoluntary: Firm initiated
Z-0997-2022Class IITerminatedVoluntary: Firm initiated