Z-1034-2022Class II
Ultrasonic Gastrovideoscope, GF-UC140P-AL5. Intended for endoscopic real-time ultrasound imaging
OLYMPUS Corporation of the Americas / initiated 2022-03-18 / Open, Classified / root cause: design
Olympus Corporation of the Americas began a recall of Ultrasonic Gastrovideoscope, GF-UC140P-AL5. Intended for endoscopic real-time ultrasound imaging on . FDA classified it as Class II and gives the reason as "GF-UC140P-AL5 is no longer a compatible endoscope for reprocessing in the OER-Pro and presents an infection control risk to patients". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1034-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- Ultrasonic Gastrovideoscope, GF-UC140P-AL5. Intended for endoscopic real-time ultrasound imaging.
- Product code
- ITX Transducer, Ultrasonic, Diagnostic
- Reason for recall
- GF-UC140P-AL5 is no longer a compatible endoscope for reprocessing in the OER-Pro and presents an infection control risk to patients
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Olympus issued URGENT: MEDICAL DEVICE CORRECTIVE ACTION on 3/18/22. Letter states reason for recall, health risk and action to take: 1. Inspect your inventory for the referenced devices and identify any device with the model number specified above. Please check all areas of the hospital to determine if any of these devices remain in inventory. The model number can be found on the device as illustrated in the following picture. 2. If you are using the OER-Pro for high level disinfection of the GF-UC140P-AL5, please transition to an OER-Elite or another AER validated for GF-UC140P-AL5 reprocessing as soon as possible. 3. When manually cleaning the GF-UC140P-AL5 carefully follow the instructions in the IFU, vigilantly cleaning all channels and the forceps elevator. Ensure that you are inspecting the GFUC140P-AL5 prior to each use and after each use according to the instructions. 4. Ensure all personnel understand that the GF-UC140P-AL5 is no longer a compatible endoscope for reprocessing in the OER-Pro and further that such personal are completely knowledgeable and thoroughly trained on the Instructions for Use applicable to the GF-UC140P-AL5 device 5. Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number 0406 and provide your contact information as indicated in the portal. If you require additional information about the GF-UC140P-AL5 reprocessing, you can obtain additional information as follows: You can contact our Technical Assistance Center (TAC) at 1-800-848-9024, option 1 and the TAC department can assist you in answering questions. OMSC requests you to report any patient injuries, including infections or persistent microbial colonization associated with any OMSC endoscope. Call our Technical Assistance Center (TAC) at 1- 800-848-9024, option 1 to report complaints.
- Quantity in commerce
- 731 units
- Distribution
- Nationwide
- Product codes and lots
- All serial numbers. GTIN: 04953170356322
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1034-2022 | Class II | Ongoing | Voluntary: Firm initiated |