Z-0886-2022Class II

Azurion systems with software release R1.x

Philips Medical Systems Nederland B.V. / initiated 2021-05-26 / Open, Classified / root cause: software

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. began a recall of Azurion systems with software release R1.x on . FDA classified it as Class II and gives the reason as "In the Azurion system, the user can add a new study to a patient by selecting the option Add Study . The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0886-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips Medical Systems Nederland B.V.
Product
Azurion systems with software release R1.x
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
In the Azurion system, the user can add a new study to a patient by selecting the option Add Study . The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a software defect, when the study is initiated by pressing Start Procedure , the Patient Type changes inadvertently to a Patient Type different than the one selected as shown in the Table below. Patient type is one of the factors involved in the dose control process. The incorrect patient type changes the technique factors to be used by the system without notification to the user.
Root cause tag
software
FDA root cause
Software design
Action
Philips notified customers of the issue via Electronic Product Radiation Defect notification letter dated March 4, 2022. The letter identified the affected product, problem and actions to be taken. Customers were informed that they will be contacted by a Philips Field Service Engineers to schedule the software update at no cost. For questions contact your local Philips representative: 1-800-722-9377.
Quantity in commerce
722 systems
Distribution
US Nationwide and Worldwide Distribution
Product codes and lots
software release R1.x

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0886-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0886-2022Class IIOngoingFDA Mandated