Z-1066-2022Class II
11274BCU1 11274BCUK1 Flexible Cystoscope Z18449US-BD (08-2018)
Karl Storz Endoscopy-America, Inc. / initiated 2022-04-01 / Open, Classified
Karl Storz Endoscopy began a recall of 11274BCU1 11274BCUK1 Flexible Cystoscope Z18449US-BD (08-2018) on . FDA classified it as Class II and gives the reason as "Failure to achieve the expected six-log reduction in microorganisms following the disinfection process". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1066-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Karl Storz Endoscopy-America, Inc.
- Product
- 11274BCU1 11274BCUK1 Flexible Cystoscope Z18449US-BD (08-2018)
- Product code
- FAJ Cystoscope And Accessories, Flexible/Rigid
- Reason for recall
- Failure to achieve the expected six-log reduction in microorganisms following the disinfection process.
- FDA root cause
- Under Investigation by firm
- Action
- On 04/01 and 6/2022, the firm emailed an "Urgent Medical Device Recall" letter to customers and followed up with a hardcopy letter on 04/07-08/2022 via their Third Party contractor informing customers of a labeling update to correct the instructions for use for certain flexible urological endoscopes and advise that high-level disinfection should not be used as a method of reprocessing. Supplemental validation testing of the efficacy of the manual high-level disinfection process using CIDEX OPA was performed. Testing showed that required efficacy level of disinfection using CIDEX OPA was not achieved. To the extent included in the current applicable instructions for use, the following methods are being removed as reprocessing methods from the instructions for use of the affected endoscopes. +Manual High-Level Disinfection: -Cidex-OPA -Revital-Ox" RESERT (e.g., 2.0% Accelerated Hydrogen Peroxide solution). +Automated Endoscope Reprocessor: -Medivators AER -Reliance EPS -EvoTech ECR Customer Actions: 1. Immediately discontinue use of all high-level disinfection methods for reprocessing affected endoscopes. 2. The affected endoscopes should be sterilized after each use by one of the sterilization methods recommended in the instructions for use specific to each endoscope. 3. If your facility does not have access to a sterilization method recommended in the applicable instructions for use, immediately discontinue use of the affected endoscopes and return products to KARL STORZ. 4. Ensure all relevant personnel in your organization are aware of recall. 5. Updated instructions for use will be available at https://spwebspace.karlstorz.com/sites/HLDCorrection/SitePages/Home.aspx . Once the updated instructions for use are available, discard any prior versions of the instructions for use. Questions: Telephone: 1-888-352-9616 Fax: 1-888-912-7088 E-Mail: [email protected]
- Quantity in commerce
- All Serial Numbers manufactured/distributed since January 2018.
- Distribution
- U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA,WI, WV, and WY. O.U.S.: Mexico, Lebanon, and South Korea
- Product codes and lots
- 11274BCU1 11274BCUK1 Flexible Cystoscope Z18449US-BD (08-2018) UDI Code: 4048551408155 All Serial Numbers manufactured/distributed since January 2018.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1061-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1062-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1063-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1064-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1065-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1066-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1067-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1068-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1069-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1070-2022 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.