Getinge 9100-Series Washer Disinfector models: 9120E, 9125E, 9127E, 9128E, 9122EW, 9128EW
Getinge Usa Sales Inc began a recall of Getinge 9100-Series Washer Disinfector models: 9120E, 9125E, 9127E, 9128E, 9122EW, 9128EW on . FDA classified it as Class II and gives the reason as "Inadequate documentation verifying whether device installation has been executed in accordance with the manufacturer's requirements. This can lead to installation failure modes and hazardous situations". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1078-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Maquet Cardiovascular Us Sales, LLC
- Product
- Getinge 9100-Series Washer Disinfector models: 9120E, 9125E, 9127E, 9128E, 9122EW, 9128EW.
- Product code
- MEC Disinfector, Medical Devices
- Reason for recall
- Inadequate documentation verifying whether device installation has been executed in accordance with the manufacturer's requirements. This can lead to installation failure modes and hazardous situations.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- An Urgent Medical Device Notification was issued 04/01/2022 via FedEx which includes serial numbers for affected devices. Customers are to verify they have an affected device and submit the attached Response Form. Distributors are to forward notification on to customers. Getinge service technicians will contact customers to schedule corrective actions. You can continue to use the Getinge Series 9100E-series Washer Disinfector with G1 control systems until the corrective action is performed, there are no precautionary actions necessary. Contact your Getinge representative with questions or for Technical Support call (888) 943-8872 (options 4,2,1), Monday through Friday, 8:00 a.m. - 6:00 p.m EST, or email [email protected].
- Quantity in commerce
- 127
- Distribution
- Worldwide distribution: (US) AZ, CA, CO, CT, DC, FL, GA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, PA, PR, SC, SD, TX, VA, WA, WV. (OUS) Australia, Austria, Belgium, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Italy, Japan, Kuwait, Libya, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Poland, Portugal, Qatar, Saudi Arabia, Serbia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam.
- Product codes and lots
- UDI Codes: [07340153710023, 07340153710047, 07340153710054, 07340153710061, 9122EW 07340153710030, 07340153710078.] Serial Numbers: WAA088008, WAA088276, WAA089054, WAA089188, WAA089511, WAA091312, WAA092134, WAA097871, WAA098978, WAA087395, WAA087675, WAA087676, WAA088563, WAA088677, WAA088684, WAA088798, WAA088973, WAA090326, WAA090827, WAA090829, WAA091562, WAA092656, WAA092905, WAA093809, WAA094122, WAA094447, WAA095653, WAA095654, WAA095655, WAA096084, WAA096723, WAA096820, WAA099698, WAA100249, WAA100596, WAA100828, WAA102416, WAA102300, WAA102301, WAA086373, WAA088830, WAA089854, WAA089735, WAA089736, WAA090141, WAA090273, WAA090274, WAA091427, WAA093786, WAA094342, WAA095013, WAA095014, WAA096724, WAA098589, WAA099164, WAA098725, WAA100514, WAA100695, WAA100704, WAA100832, WAA101375, WAA100834, WAA101893, WAA076270, WAA085559, WAA086376, WAA086914, WAA086915, WAA088007, WAA088273, WAA088464, WAA088465, WAA087754, WAA088678, WAA088685, WAA088871, WAA088998, WAA089258, WAA089356, WAA089446, WAA089970, WAA089969, WAA090495, WAA090615, WAA090746, WAA091561, WAA091184, WAA091560, WAA091563, WAA091564, WAA091869, WAA091870, WAA091873, WAA091868, WAA092136, WAA091872, WAA092904, WAA093646, WAA093882, WAA093967, WAA094123, WAA094202, WAA094284, WAA094285, WAA094286, WAA094617, WAA094601, WAA095421, WAA096425, WAA096427, WAA096428, WAA097298, WAA098056, WAA099694, WAA099450, WAA099451, WAA099695, WAA099697, WAA100127, WAA100826, WAA100827, WAA100830, WAA101342, WAA101894, WAA103173, WAA088274, WAA093968
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1078-2022 | Class II | Ongoing | Voluntary: Firm initiated |