Z-1095-2022Class I
Dragonfly OpStar Imaging Catheter, REF 1014651/1014652, 0.068 in, 0.014 in, 135 cm, STERILE EO
Abbott Vascular / initiated 2022-03-30 / Open, Classified
Abbott Vascular began a recall of Dragonfly OpStar Imaging Catheter, REF 1014651/1014652, 0.068 in, 0.014 in, 135 cm, STERILE EO on . FDA classified it as Class I and gives the reason as "There is a potential that the proximal marker may separate from the imaging catheter". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1095-2022
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abbott Vascular
- Product
- Dragonfly OpStar Imaging Catheter, REF 1014651/1014652, 0.068 in, 0.014 in, 135 cm, STERILE EO
- Product code
- DQO Catheter, Intravascular, Diagnostic
- Reason for recall
- There is a potential that the proximal marker may separate from the imaging catheter
- FDA root cause
- Under Investigation by firm
- Action
- On April 11, 2022, Abbott issued a "Urgent: Medical Device Recall" Notification to all affected consignees. On June 6, 2022, Abbott issued a updated Urgent Medical Device Recall notification to provide new information concerning the recall product. Abbott ask consignees to take the following actions: 1. Immediately stop using devices from affected lots 2. Return all unused affected devices to Abbott (if not already completed per previous communication) 3. Review this updated notification letter 4. Complete and return the provided Effectiveness Check Form 5 Share this notification with relevant personnel in your organization. 6If you have further distributed/transferred the affected products, notify those customers. 7. Report any occurrence of product performance issues or patient adverse event to Abbott. 8. Reports may also be sent to the FDA's MedWatch Adverse Event Reporting program by completing Form FDA 3500 online at www.fda.gov or calling 1-800-FDA-1088 or submit by mail or fax to 1-800-FDA-0178
- Quantity in commerce
- 14,544 UNITS
- Distribution
- US:AL AR AZ CA CO CT DC FL GA HI IA IL IN KS KY MA MD ME MI MN MO MT NC NE NH NJ NM NV NY OH OK OR PA PR RI SC TN TX UT VA VT WA WI, PR OUS: AE, AT, AW, BE, CA, CH, DE, DK, IT, JP, LU, NL, NO, NZ, PL, EE, ES, FI, FR, GB, HK, HU, IE, PT, RO, SE, SG, SI, TH
- Product codes and lots
- Model 1014651 - GTIN: (01)05415067031129 Lot #: 8010027 8147191 8010043 8153815 8010047 8153816 8010052 8211505 8127930, 8211506, 8127931, 8211507, 8127932, 8220671, 8127934, 8220672, 8131366, 8220673, 8131367, 8220675, 8131369, 8257085, 8131370, 8257087, 8147186, 8274128 8147188, 8294077 Model 1014652 - GTIN: (01) 05415067031112 Lot #:8111644, 8192796, 8131360, 8211508, 8131361, 8211509, 8131362, 8211510,8131363 8211511, 8131364, 8211512, 8131365, 8211513, 8157505, 8211514, 8157506, 8211515 8184979, 8211516, 8184980, 8211517, 8184981, 8211518, 8184982, 8211519, 8185100 8211520, 8185101, 8211521, 8192782, 8220680, 8192783, 8220681, 8192784, 8220684, 8192786, 8220685, 8192787, 8220686, 8192788, 8256952, 8192789, 8256953, 8192791 8256954, 8192793, 8274133, 8192795, 8294003