Z-1126-2022Class II

(1) Gibeck Humid-Vent 2S Flex DC, Product #17731, sterile; (2) Gibeck Humid-Vent 2S Flex DC, Product #17732...

Teleflex Medical / initiated 2022-04-11 / Open, Classified / root cause: design

TELEFLEX MEDICAL INC began a recall of (1) Gibeck Humid-Vent 2S Flex DC, Product #17731, sterile; (2) Gibeck Humid-Vent 2S Flex DC, Product #17732, non-sterile; (3) Gibeck Humid-Vent Filter Compact... on . FDA classified it as Class II and gives the reason as "Reports received of torn corrugated breathing tubes". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1126-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Teleflex Medical
Product
(1) Gibeck Humid-Vent 2S Flex DC, Product #17731, sterile; (2) Gibeck Humid-Vent 2S Flex DC, Product #17732, non-sterile; (3) Gibeck Humid-Vent Filter Compact S with Smooth-Flo/Swivel, Product #19061, sterile; (4) Gibeck Humid-Vent Filter Compact S with Smooth-Flo/Swivel, Product #19062, non-sterile; and (5) Gibeck Humid-Vent Filter Compact S with Flex Tube, Product #19812, non-sterile.
Product code
BYD Condenser, Heat And Moisture (Artificial Nose)
Reason for recall
Reports received of torn corrugated breathing tubes.
Root cause tag
design
FDA root cause
Device Design
Action
The recalling firm issued letters dated April 2022 on 4/11/2022 via FedEx 2-day mail. The letter described the issue, the immediate actions required including the hazards involved, and provided specific actions to be taken and instructions to be followed by the medical facilities or the distributors. For the medical facilities, they are to cease use and distribution of the impacted product and immediately quarantine it. If the product is in inventory, indicate the information on the Acknowledgment Form and FAX it to Customer Service or email it to the firm. A customer service representative will contact the medical facility with a Return Goods Authorization number and provide instructions for return of the product to Teleflex. The Acknowledgment Form should also be returned if no product is in inventory. For distributors, they are to provide the recall letter to all customers who have received the product. Each customer is required to complete and return the Acknowledgement Form. The distributor is also supposed to check their inventory, cease distribution and quarantine the affected product. The Acknowledgment Form is to be completed and returned for product in their inventory as well as if they have no inventory. If the distributor has distributed product outside of the country, they are to notify Teleflex Customer Service. Additionally, the recall letter says that this notice should be passed onto all persons who need to be aware within their organization or to any organizations where the potentially affected devices have been transferred. This would include users, clinicians, risk managers, supply chain/distribution centers, etc. in the circulation of the notice.
Quantity in commerce
59,581 eaches
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NJ, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, WI, and WV. The countries of Austria, Australia, Belgium, Canada, Chile, Czech Republic, Finland, France, Germany, Greece, Hong Kong, Hungary, Korea, Malaysia, Netherlands, Panama, Poland, Slovakia, Spain, and Switzerland.
Product codes and lots
Batch and UDI numbers: (1) Product 17731: KMH21C0130 (01)04026704347919(17)260228(10)KMH21C0130; KMH21C0162 (01)04026704347919(17)260228(10)KMH21C0162; KMH21C0262 (01)04026704347919(17)260228(10)KMH21C0262; KMH21C0265 (01)04026704347919(17)260228(10)KMH21C0265; KMH21E0077 (01)04026704347919(17)260428(10)KMH21E0077; KMH21F0231 (01)04026704347919(17)260528(10)KMH21F0231; KMH21F0276 (01)04026704347919(17)260528(10)KMH21F0276; KMH21G0428 (01)04026704347919(17)260628(10)KMH21G0428; KMH21H0142 (01)04026704347919(17)260728(10)KMH21H0142; KMH21L0020 (01)04026704347919(17)261028(10)KMH21L0020; and KMH22A0068 (01)04026704347919(17)261228(10)KMH22A0068. (2) Product 17732: KMZ21G0733 (01)04026704347926(17)260628(10)KMZ21G0733; KMZ21G0742 (01)04026704347926(17)260628(10)KMZ21G0742; KMZ21G0989 (01)04026704347926(17)260628(10)KMZ21G0989; KMZ21G1049 (01)04026704347926(17)260628(10)KMZ21G1049; KMZ21K0159 (01)04026704347926(17)260928(10)KMZ21K0159; KMZ21K0160 (01)04026704347926(17)260928(10)KMZ21K0160; KMZ21K0275 (01)04026704347926(17)260928(10)KMZ21K0275; KMZ21K0451 (01)04026704347926(17)260928(10)KMZ21K0451; KMZ21L0103 (01)04026704347926(17)261028(10)KMZ21L0103; KMZ21L0602 (01)04026704347926(17)261028(10)KMZ21L0602; KMZ21C0125 (01)04026704347926(17)260228(10)KMZ21C0125; KMZ21C0235 (01)04026704347926(17)260228(10)KMZ21C0235; KMZ21C0300 (01)04026704347926(17)260228(10)KMZ21C0300; KMZ21C0347 (01)04026704347926(17)260228(10)KMZ21C0347; KMZ21C0351 (01)04026704347926(17)260228(10)KMZ21C0351; KMZ21C0475 (01)04026704347926(17)260228(10)KMZ21C0475; KMZ21C0609 (01)04026704347926(17)260228(10)KMZ21C0609; KMZ21C0851 (01)04026704347926(17)260228(10)KMZ21C0851; KMZ21C0863 (01)04026704347926(17)260228(10)KMZ21C0863; KMZ21C1154 (01)04026704347926(17)260228(10)KMZ21C1154; KMZ21C1342 (01)04026704347926(17)260228(10)KMZ21C1342; KMZ21D0016 (01)04026704347926(17)260328(10)KMZ21D0016; KMZ21D0200 (01)04026704347926(17)260328(10)KMZ21D0200; KMZ21D0311 (01)04026704347926(17)260328(10)KMZ21D0311; KMZ21D0357 (01)04026704347926(17)260328(10)KMZ21D0357; KMZ21D0404 (01)04026704347926(17)260328(10)KMZ21D0404; KMZ21D0481 (01)04026704347926(17)260328(10)KMZ21D0481; KMZ21D0510 (01)04026704347926(17)260328(10)KMZ21D0510; KMZ21E0096 (01)04026704347926(17)260428(10)KMZ21E0096; KMZ21F0475 (01)04026704347926(17)260528(10)KMZ21F0475; KMZ21F0483 (01)04026704347926(17)260528(10)KMZ21F0483; KMZ21F0484 (01)04026704347926(17)260528(10)KMZ21F0484; and KMZ21F0873 (01)04026704347926(17)260528(10)KMZ21F0873. (3) Product 19061: KMH21F0418 (01)04026704388387(17)260528(10)KMH21F0418; KMH21G0357 (01)04026704388387(17)260628(10)KMH21G0357; KMH21H0060 (01)04026704388387(17)260728(10)KMH21H0060; KMH21H0062 (01)04026704388387(17)260728(10)KMH21H0062; KMH21H0252 (01)04026704388387(17)260728(10)KMH21H0252; KMH21J0021 (01)04026704388387(17)260828(10)KMH21J0021; KMH21J0260 (01)04026704388387(17)260828(10)KMH21J0260; KMH21K0159 (01)04026704388387(17)260928(10)KMH21K0159; KMH21K0375 (01)04026704388387(17)260928(10)KMH21K0375; KMH21L0139 (01)04026704388387(17)261028(10)KMH21L0139; KMH21C0188 (01)04026704388387(17)260228(10)KMH21C0188; KMH21C0191 (01)04026704388387(17)260228(10)KMH21C0191; KMH21C0194 (01)04026704388387(17)260228(10)KMH21C0194; KMH21C0280 (01)04026704388387(17)260228(10)KMH21C0280; KMH21C0388 (01)04026704388387(17)260228(10)KMH21C0388; KMH21D0094 (01)04026704388387(17)260328(10)KMH21D0094; KMH21D0289 (01)04026704388387(17)260328(10)KMH21D0289; KMH21D0290 (01)04026704388387(17)260328(10)KMH21D0290; KMH21E0240 (01)04026704388387(17)260428(10)KMH21E0240; and KMH21F0155 (01)04026704388387(17)260528(10)KMH21F0155. (4) Product 19062: KMZ21C0425 (01)04026704388394(17)260228(10)KMZ21C0425; KMZ21C0923 (01)04026704388394(17)260228(10)KMZ21C0923; and KMZ21F0379 (01)04026704388394(17)260528(10)KMZ21F0379. (5) Product 19812: KMZ21G0681 (01)04026704388721(17)260628(10)KMZ21G0681; KMZ21G0710 (01)04026704388721(17)260628(10)KMZ21G0710; KMZ21K0459 (01)04026704388721(17)260928(10)KMZ21K0459; KMZ21C0141 (01)04026704388721(17)260228(10)KMZ21C0141; KMZ21C0948 (01)04026704388721(17)260228(10)KMZ21C0948; KMZ21F0361 (01)04026704388721(17)260528(10)KMZ21F0361; and KMZ21F0627 (01)04026704388721(17)260528(10)KMZ21F0627.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1126-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1126-2022Class IIOngoingVoluntary: Firm initiated
Z-1127-2022Class IIOngoingVoluntary: Firm initiated