Z-1092-2022Class II
Verigene Processor SP, Models 10-0000-07/10-0000-07R a component of the Verigene system
Luminex Corporation / initiated 2022-04-05 / Open, Classified
Luminex Corporation began a recall of Verigene Processor SP, Models 10-0000-07/10-0000-07R a component of the Verigene system on . FDA classified it as Class II and gives the reason as "Heater within the Verigene System may be outside of established temperature limits for assays which could result in false negative results". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1092-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Luminex Corporation
- Product
- Verigene Processor SP, Models 10-0000-07/10-0000-07R a component of the Verigene system
- Product code
- OOI Real Time Nucleic Acid Amplification System
- Reason for recall
- Heater within the Verigene System may be outside of established temperature limits for assays which could result in false negative results.
- FDA root cause
- Under Investigation by firm
- Action
- On 04/21/22 Luminex issued a "Urgent Medical Device Recall" Notification to affected consignees. On 6/16/22 and 7/14/22 Luminex expanded the recall to include additional serial numbers. Luminex is asking consignees to take the following actions: 1. Continue use of your VERIGENE Processor SP as service has been completed by a Luminex Field Service Engineer and the Processor SP s heaters are now within specification. 2. Perform a retrospective review of results for VERIGENE Gram-Positive Blood Culture (BC-GP) and VERIGENE Gram-Negative Blood Culture (BC-GN) Nucleic Acid Tests from 18 months prior to the most recent service visit. Identify any results that were not consistent with other clinical signs and symptoms. Promptly inform Luminex of any previously unreported incidences of false results. 3. Distribute the notification to any individuals within your organization who need to be made aware. 4. Please reach out to Global Support Services if a copy of the service record is required. 5. Please complete the Acknowledgement & Receipt form and return within 10 business days.
- Quantity in commerce
- 24 units
- Distribution
- US:OH, ME, PA, FL, IL, NE, IN, TX, GA, WA, MI, CO, AZ, OH, KY, CA, VT OUS: None
- Product codes and lots
- Serial Numbers for 10-0000-07: 17340003 19044007 20133004 18128002 18214004 12061048 18191010 13193254 13051068 21029902 16277059 19233005 11250096 14069051 18255003 *18039010 *20212006 *19317004 *19206009 *18057008 Serial Numbers added 6.23.22 for 10-0000-07R: 18191010 13051068 14069051 *12198198 *12278344 *14322267 *12271338 Serial Numbers added 7.14.22 for 10-0000-07R: 18226007 and 16242043 Serial Numbers added 7.14.22 for 10-0000-07: 17062030, 18235008 and 19192006
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1092-2022 | Class II | Ongoing | Voluntary: Firm initiated |