Z-1092-2022Class II

Verigene Processor SP, Models 10-0000-07/10-0000-07R a component of the Verigene system

Luminex Corporation / initiated 2022-04-05 / Open, Classified

Luminex Corporation began a recall of Verigene Processor SP, Models 10-0000-07/10-0000-07R a component of the Verigene system on . FDA classified it as Class II and gives the reason as "Heater within the Verigene System may be outside of established temperature limits for assays which could result in false negative results". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1092-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Luminex Corporation
Product
Verigene Processor SP, Models 10-0000-07/10-0000-07R a component of the Verigene system
Product code
OOI Real Time Nucleic Acid Amplification System
Reason for recall
Heater within the Verigene System may be outside of established temperature limits for assays which could result in false negative results.
FDA root cause
Under Investigation by firm
Action
On 04/21/22 Luminex issued a "Urgent Medical Device Recall" Notification to affected consignees. On 6/16/22 and 7/14/22 Luminex expanded the recall to include additional serial numbers. Luminex is asking consignees to take the following actions: 1. Continue use of your VERIGENE Processor SP as service has been completed by a Luminex Field Service Engineer and the Processor SP s heaters are now within specification. 2. Perform a retrospective review of results for VERIGENE Gram-Positive Blood Culture (BC-GP) and VERIGENE Gram-Negative Blood Culture (BC-GN) Nucleic Acid Tests from 18 months prior to the most recent service visit. Identify any results that were not consistent with other clinical signs and symptoms. Promptly inform Luminex of any previously unreported incidences of false results. 3. Distribute the notification to any individuals within your organization who need to be made aware. 4. Please reach out to Global Support Services if a copy of the service record is required. 5. Please complete the Acknowledgement & Receipt form and return within 10 business days.
Quantity in commerce
24 units
Distribution
US:OH, ME, PA, FL, IL, NE, IN, TX, GA, WA, MI, CO, AZ, OH, KY, CA, VT OUS: None
Product codes and lots
Serial Numbers for 10-0000-07: 17340003 19044007 20133004 18128002 18214004 12061048 18191010 13193254 13051068 21029902 16277059 19233005 11250096 14069051 18255003 *18039010 *20212006 *19317004 *19206009 *18057008 Serial Numbers added 6.23.22 for 10-0000-07R: 18191010 13051068 14069051 *12198198 *12278344 *14322267 *12271338 Serial Numbers added 7.14.22 for 10-0000-07R: 18226007 and 16242043 Serial Numbers added 7.14.22 for 10-0000-07: 17062030, 18235008 and 19192006

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1092-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1092-2022Class IIOngoingVoluntary: Firm initiated