VITROS Immunodiagnostic Products FSH Reagent Pack- IVD 1931922
Ortho-Clinical Diagnostics, Inc began a recall of VITROS Immunodiagnostic Products FSH Reagent Pack- IVD 1931922 on . FDA classified it as Class II and gives the reason as "Falsely high or delayed patient and QC results may occur due to low end imprecision. VITROS Immunodiagnostic Products FSH, LH and Prol Reagents Low End Imprecision. Low-level quality control (QC) and patient samples at...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1112-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS Immunodiagnostic Products FSH Reagent Pack- IVD 1931922
- Product code
- CGJ Radioimmunoassay, Follicle-Stimulating Hormone
- Reason for recall
- Falsely high or delayed patient and QC results may occur due to low end imprecision. VITROS Immunodiagnostic Products FSH, LH and Prol Reagents Low End Imprecision. Low-level quality control (QC) and patient samples at the low end of the Measuring (Reportable) Range when using VITROS FSH, LH and Prol Reagent. Calibration failures, driven by imprecision observed with Calibrator Level 1.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Ortho CLinical Diagnostics issued "URGENT PRODUCT CORRECTION NOTIFICATION" dated 3/25/22 to consignees who had been shipped VITROS FSH, LH or Prol in the previous 18 months were notified (CL2022-069) of the issue and instructed as follows: -Run daily low-level QC in duplicate to confirm acceptable performance. -If their laboratory is experiencing low end imprecision with VITROS FSH, LH or Prol Reagent, follow one of the suggested testing algorithms as described in the enclosure. o Note: If their laboratory is not experiencing low end imprecision, continue testing patient samples as usual (single replicate). -Until further notice, all future lots of the products will also be affected. -On the enclosed Credit Form, track the additional quantities of VITROS FSH, LH and/or Prol tests used due to the testing protocol. Periodically submit the form to Ortho. Ortho will credit their account. If you have further questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.
- Quantity in commerce
- 2502 total: 2154 US; 348 OUS
- Distribution
- Worldwide Distribution: US (Nationwide) OUS (Foreign): Australia Bermuda Brazil Canada L3R 4G5 Chile China Colombia Denmark France Germany India Italy Japan Mexico Netherlands Norway Portugal Russia Singapore Spain Sweden United Kingdom
- Product codes and lots
- Until further notice, future lots will also be affected. UDI: 10758750000302 Affected Lot Number/ Expiration Date: 3100 01-MAR-2022; 3111 25-MAR-2022; 3120 04-APR-2022; 3130 25-APR-2022; 3140 10-MAY-2022; 3145 10-MAY-2022; 3150 30-MAY-2022; 3160 09-JUN-2022; 3170 22-JUN-2022; 3180 02-AUG-2022; 3190 16-AUG-2022; 3200 06-SEP-2022; 3210 12-SEP-2022.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1112-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1113-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1114-2022 | Class II | Ongoing | Voluntary: Firm initiated |