Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1244-2022Class II

V8 Immunodisplacement Kit REF 1803

Helena Laboratories, Corp. / initiated 2022-04-06 / Open, Classified

Helena Laboratories, Corp. began a recall of V8 Immunodisplacement Kit REF 1803 on . FDA classified it as Class II and gives the reason as "Due to microbial growth causing interference with interpretation". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1244-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Helena Laboratories, Corp.
Product
V8 Immunodisplacement Kit REF 1803
Product code
CFF Immunoelectrophoretic, Immunoglobulins, (G, A, M)
Reason for recall
Due to microbial growth causing interference with interpretation.
FDA root cause
Unknown/Undetermined by firm
Action
On 04/06/2022, Helena Laboratories emailed an "URGENT MEDICAL DEVICE RECALL" Letter to customers informing them that the firm is recalling V8 Immunodisplacement Kit Lots 3-21-1803, and 1-22-1803 due to visible particulate contamination in some vials of a component within the kit - the IgG antisera. Use of this product may result in an artifact anodal to albumin potentially interfering with interpretation of the IgG pattern. Customers are instructed to immediately evaluate their inventory for product subject to recall, and quarantine accordingly. All unused bottles in the field will be removed from the field, and replaced with an unaffected lot. For questions or further assistance, email or call Dr. Jessica Jones Hanka at [email protected] or 1-800-231-5663 extension 1600.
Quantity in commerce
51 kits
Distribution
U.S.: FL, MO, OH, and TX O.U.S.: Canada, Uruguay, and Vietnam,
Product codes and lots
Model Number: 1803 UDI Codes: +M52518030/SS73-21-1803/14D20230831Z +M52518030/SS71-21-1803/14D20230830X Lot 3-21-1803 Lot 1-22-1803

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1244-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1244-2022Class IIOngoingVoluntary: Firm initiated