Z-1244-2022Class II
V8 Immunodisplacement Kit REF 1803
Helena Laboratories, Corp. / initiated 2022-04-06 / Open, Classified
Helena Laboratories, Corp. began a recall of V8 Immunodisplacement Kit REF 1803 on . FDA classified it as Class II and gives the reason as "Due to microbial growth causing interference with interpretation". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1244-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Helena Laboratories, Corp.
- Product
- V8 Immunodisplacement Kit REF 1803
- Product code
- CFF Immunoelectrophoretic, Immunoglobulins, (G, A, M)
- Reason for recall
- Due to microbial growth causing interference with interpretation.
- FDA root cause
- Unknown/Undetermined by firm
- Action
- On 04/06/2022, Helena Laboratories emailed an "URGENT MEDICAL DEVICE RECALL" Letter to customers informing them that the firm is recalling V8 Immunodisplacement Kit Lots 3-21-1803, and 1-22-1803 due to visible particulate contamination in some vials of a component within the kit - the IgG antisera. Use of this product may result in an artifact anodal to albumin potentially interfering with interpretation of the IgG pattern. Customers are instructed to immediately evaluate their inventory for product subject to recall, and quarantine accordingly. All unused bottles in the field will be removed from the field, and replaced with an unaffected lot. For questions or further assistance, email or call Dr. Jessica Jones Hanka at [email protected] or 1-800-231-5663 extension 1600.
- Quantity in commerce
- 51 kits
- Distribution
- U.S.: FL, MO, OH, and TX O.U.S.: Canada, Uruguay, and Vietnam,
- Product codes and lots
- Model Number: 1803 UDI Codes: +M52518030/SS73-21-1803/14D20230831Z +M52518030/SS71-21-1803/14D20230830X Lot 3-21-1803 Lot 1-22-1803
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1244-2022 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.