Z-1138-2022Class I

In-Line ventilator adaptor

Baxter Healthcare Corporation / initiated 2022-04-26 / Terminated / root cause: design

Baxter Healthcare Corporation began a recall of In-Line ventilator adaptor on . FDA classified it as Class I and gives the reason as "There is a potential of reduced oxygenation or pneumothorax/barotrauma to occur when the Volara system is used with in-line ventilator in home care environment". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1138-2022
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
In-Line ventilator adaptor
Product code
NHJ Device, Positive Pressure Breathing, Intermittent
Reason for recall
There is a potential of reduced oxygenation or pneumothorax/barotrauma to occur when the Volara system is used with in-line ventilator in home care environment.
Root cause tag
design
FDA root cause
Device Design
Action
The recall notification was sent to consignees on 04/26/2022 via FedEx. The customer was instructed to sign acknowledgement form to confirm receipt and return to Hillrom using pre-paid envelope.
Quantity in commerce
259 units
Distribution
US Nationwide distribution in the states of TX, FL, VA, KY, CA, AZ, WI, IL, NC, OH, MI, MS, MO, MN, AL, TN, SD, OK, NV, NJ, IN, CO, GA, WA, PA, IA, NE, NM, SC, ND, KS, WV, OR.
Product codes and lots
M08473 OPTIMUS OLE AC PAT.CIRCUIT KIT; UDI-DI: 00887761981492. Manufacturing dates February 28, 2020 to present.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1138-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1138-2022Class ITerminatedVoluntary: Firm initiated
Z-1139-2022Class ITerminatedVoluntary: Firm initiated