Z-1116-2022Class II
Olympus Uretero-reno videoscope, Model No. URF-V2
OLYMPUS Corporation of the Americas / initiated 2022-03-25 / Open, Classified
Olympus Corporation of the Americas began a recall of Olympus Uretero-reno videoscope, Model No. URF-V2 on . FDA classified it as Class II and gives the reason as "The bending section of the URF-V2 may crack, resulting in an abnormal shape or abnormal angulation of the bending section". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1116-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- Olympus Uretero-reno videoscope, Model No. URF-V2
- Product code
- NWB Endoscope, Accessories, Narrow Band Spectrum
- Reason for recall
- The bending section of the URF-V2 may crack, resulting in an abnormal shape or abnormal angulation of the bending section.
- FDA root cause
- Process change control
- Action
- On March 25, 2022, the firm contacted the affected consignee to inform them that certain manufacturing changes by a supplier led to the possibility that a part within the bending section may crack. Olympus repaired the affected URF-V2 unit free of charge.
- Quantity in commerce
- 1
- Distribution
- US Nationwide distribution in the state of AZ.
- Product codes and lots
- 1 unit, serial number 2825505F UDI 04953170343582
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1116-2022 | Class II | Ongoing | Voluntary: Firm initiated |