Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1116-2022Class II

Olympus Uretero-reno videoscope, Model No. URF-V2

OLYMPUS Corporation of the Americas / initiated 2022-03-25 / Open, Classified

Olympus Corporation of the Americas began a recall of Olympus Uretero-reno videoscope, Model No. URF-V2 on . FDA classified it as Class II and gives the reason as "The bending section of the URF-V2 may crack, resulting in an abnormal shape or abnormal angulation of the bending section". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1116-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
Olympus Uretero-reno videoscope, Model No. URF-V2
Product code
NWB Endoscope, Accessories, Narrow Band Spectrum
Reason for recall
The bending section of the URF-V2 may crack, resulting in an abnormal shape or abnormal angulation of the bending section.
FDA root cause
Process change control
Action
On March 25, 2022, the firm contacted the affected consignee to inform them that certain manufacturing changes by a supplier led to the possibility that a part within the bending section may crack. Olympus repaired the affected URF-V2 unit free of charge.
Quantity in commerce
1
Distribution
US Nationwide distribution in the state of AZ.
Product codes and lots
1 unit, serial number 2825505F UDI 04953170343582

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1116-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1116-2022Class IIOngoingVoluntary: Firm initiated