Z-1028-2022Class II
Evans Wedge Disposable Trials, Large Footprint, 8mm Thickness (Model: 6101-22200818). Orthopedic manual...
restor3d, Inc. / initiated 2022-03-30 / Terminated
restor3d Inc. began a recall of Evans Wedge Disposable Trials, Large Footprint, 8mm Thickness (Model: 6101-22200818). Orthopedic manual surgical instrument on . FDA classified it as Class II and gives the reason as "During re-work of disposable instruments to relabel for 4-year shelf-life of disposable instruments, parts got mixed into other lots". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1028-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- restor3d, Inc.
- Product
- Evans Wedge Disposable Trials, Large Footprint, 8mm Thickness (Model: 6101-22200818). Orthopedic manual surgical instrument
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- During re-work of disposable instruments to relabel for 4-year shelf-life of disposable instruments, parts got mixed into other lots.
- FDA root cause
- Process change control
- Action
- An email was sent out to the consignees on 03/31/2022. All customers were instructed to segregate any affected product and contact Restor3d. If all product has already been returned to restor3d, the consignee is to indicate that on the enclosed response form. On 04/19/2022, the firm extended the recall to two additional lots. The recall notification was sent to the consignees via email with a request for all affected product to be returned.
- Quantity in commerce
- 9 units
- Distribution
- US Nationwide distribution in the states of KY, NC, NJ, TX, IN, VA and AL.
- Product codes and lots
- Model: 6101-22200818; Lot: 2022010035; UDI: 00840097501598
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1026-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1027-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1028-2022 | Class II | Terminated | Voluntary: Firm initiated |