Cardioblate" Gemini"-s Irrigated RF Surgical Ablation System; The device is placed in a plastic tray with a...
Medtronic Perfusion Systems began a recall of Cardioblate" Gemini"-s Irrigated RF Surgical Ablation System; The device is placed in a plastic tray with a snap on plastic lid, the tray is placed in... on . FDA classified it as Class II and gives the reason as "Firm detected an increase in complaints related to fractured jaw tips of the Cardioblate Gemini-s". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1131-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic Perfusion Systems
- Product
- Cardioblate" Gemini"-s Irrigated RF Surgical Ablation System; The device is placed in a plastic tray with a snap on plastic lid, the tray is placed in protective wrap, then into a sterile pouch, 1 device per carton, 1 carton per box.
- Product code
- OCL Surgical Device, For Cutting, Coagulation, And/Or Ablation Of Tissue, Including Cardiac Tissue
- Reason for recall
- Firm detected an increase in complaints related to fractured jaw tips of the Cardioblate Gemini-s.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- A recall notification was sent to consignees on 04/20/2022 via the United Parcel Service. The letter instructs the consignees to review the updated guidance on device usage contained in the letter, share the notice to those that need to be aware, and complete the enclosed Customer Confirmation Form and email to [email protected].
- Quantity in commerce
- 1236 units
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, ND, MA, ID, WI, KS, UT, MN, NY, NJ, GA, MO and the countries of Australia, Belgium, China, Czech Republic, France, Germany, Greece, Israel, Italy, Kazakhstan, Latvia, Netherlands, Poland, Russian Federation, Singapore, Switzerland.
- Product codes and lots
- ABLATION DEVICE 49260 PROCEDURE KIT Model 49260; GTIN: 00613994268051, Lot Numbers: 215G, 256E, 388C, 466C, 584D, 611A, 745F, 840C, 850C, 967C; GTIN: 00613994755643, Lot Numbers: 469C, 612A, 972C, 973C, GTIN: 00643169983120, Lot Numbers: 125E, 238D, 257E, 258E, 259E, 335D, 373B, 410E, 428D, 454F, 467C, 585D, 632E, 635F, 700D, 721C, 722C, 732E, 735E, 738E, 841C, 876E, 923D, 968C, 972E, 980E, 995D. CLAMP 49351 CARDIOBLATE GEMINI-S FT Model: 49351; GTIN: 00643169998117, Lot Numbers: 495F, 734E, 846D.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1131-2022 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.