Z-1131-2022Class II

Cardioblate" Gemini"-s Irrigated RF Surgical Ablation System; The device is placed in a plastic tray with a...

Medtronic Perfusion Systems / initiated 2022-04-20 / Open, Classified / root cause: labeling and UDI

Medtronic Perfusion Systems began a recall of Cardioblate" Gemini"-s Irrigated RF Surgical Ablation System; The device is placed in a plastic tray with a snap on plastic lid, the tray is placed in... on . FDA classified it as Class II and gives the reason as "Firm detected an increase in complaints related to fractured jaw tips of the Cardioblate Gemini-s". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1131-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic Perfusion Systems
Product
Cardioblate" Gemini"-s Irrigated RF Surgical Ablation System; The device is placed in a plastic tray with a snap on plastic lid, the tray is placed in protective wrap, then into a sterile pouch, 1 device per carton, 1 carton per box.
Product code
OCL Surgical Device, For Cutting, Coagulation, And/Or Ablation Of Tissue, Including Cardiac Tissue
Reason for recall
Firm detected an increase in complaints related to fractured jaw tips of the Cardioblate Gemini-s.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
A recall notification was sent to consignees on 04/20/2022 via the United Parcel Service. The letter instructs the consignees to review the updated guidance on device usage contained in the letter, share the notice to those that need to be aware, and complete the enclosed Customer Confirmation Form and email to [email protected].
Quantity in commerce
1236 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of CA, ND, MA, ID, WI, KS, UT, MN, NY, NJ, GA, MO and the countries of Australia, Belgium, China, Czech Republic, France, Germany, Greece, Israel, Italy, Kazakhstan, Latvia, Netherlands, Poland, Russian Federation, Singapore, Switzerland.
Product codes and lots
ABLATION DEVICE 49260 PROCEDURE KIT Model 49260; GTIN: 00613994268051, Lot Numbers: 215G, 256E, 388C, 466C, 584D, 611A, 745F, 840C, 850C, 967C; GTIN: 00613994755643, Lot Numbers: 469C, 612A, 972C, 973C, GTIN: 00643169983120, Lot Numbers: 125E, 238D, 257E, 258E, 259E, 335D, 373B, 410E, 428D, 454F, 467C, 585D, 632E, 635F, 700D, 721C, 722C, 732E, 735E, 738E, 841C, 876E, 923D, 968C, 972E, 980E, 995D. CLAMP 49351 CARDIOBLATE GEMINI-S FT Model: 49351; GTIN: 00643169998117, Lot Numbers: 495F, 734E, 846D.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1131-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1131-2022Class IIOngoingVoluntary: Firm initiated