Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) began a recall of The C304-HIS device features a guide wire to access the vein, a valve to reduce blood loss during the implant procedure, a deflectable catheter to introduce a... on . FDA classified it as Class II and gives the reason as "The firm's internal processes identified that certain lots may have an improperly sealed barrier that could potentially compromise sterility". As of , the recall status on record is Open, Classified, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-1206-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- The C304-HIS device features a guide wire to access the vein, a valve to reduce blood loss during the implant procedure, a deflectable catheter to introduce a transvenous device, a catheter dilator to facilitate deflectable catheter passage, and a guide catheter slitter to remove the deflectable catheter. The deflectable catheter is designed for placement of transvenous devices in or near the bundle of His. It features a deflecting distal section, controlled by the deflectable catheter handle, and an out-of plane curve on the distal tip. The body of the deflectable catheter is radiopaque for visibility on fluoroscopy.
- Product code
- DQY Catheter, Percutaneous
- Reason for recall
- The firm's internal processes identified that certain lots may have an improperly sealed barrier that could potentially compromise sterility.
- Root cause tag
- sterility
- FDA root cause
- Under Investigation by firm
- Action
- The firm sent out a recall notification starting on 04/20/2022 via UPS 2-day delivery. Customers are asked to return all unused affected product, share the notice with all those who need to be aware within their organization or with any organization where the above-listed products have been transferred, and to complete the response form.
- Quantity in commerce
- 1385 devices
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV and the countries of Belgium, Denmark, France, Germany, Ireland, Italy, Netherlands, Poland, Spain, United Arab Emirates, United Kingdom.
- Product codes and lots
- Model Number: C304-HIS; GTIN: 00763000147013; Lot Numbers: 0010948864, 0010953298, 0010958150, 0010961222, 0010971148, 0011017539, 0011020263, 0011022874, 0011025202, 0011028574, 0011031177
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1206-2022 | Class II | Ongoing | Voluntary: Firm initiated |