Z-1240-2022Class II

Hilal Embolization Microcoil, RPN MWCE-18-1.0-0-HILAL-01 (GPN G47499); RPN MWCE-18-0.5-0-HILAL-01 (GPN G47498)

Cook Incorporated / initiated 2022-05-05 / Open, Classified

Cook Incorporated began a recall of Hilal Embolization Microcoil, RPN MWCE-18-1.0-0-HILAL-01 (GPN G47499); RPN MWCE-18-0.5-0-HILAL-01 (GPN G47498) on . FDA classified it as Class II and gives the reason as "The loading cartridges included with these devices may contain a small, unintended, stainless-steel cannula". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1240-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Cook Incorporated
Product
Hilal Embolization Microcoil, RPN MWCE-18-1.0-0-HILAL-01 (GPN G47499); RPN MWCE-18-0.5-0-HILAL-01 (GPN G47498)
Product code
KRD Device, Vascular, For Promoting Embolization
Reason for recall
The loading cartridges included with these devices may contain a small, unintended, stainless-steel cannula.
FDA root cause
Under Investigation by firm
Action
On May 5, 2022, the firm distributed Urgent Medical Device Recall letters to affected customers. Customers were advised that the presence of a small stainless steel cannula inside the loading cartridge may go undetected by the user. Customers were instructed to examine their inventory and quarantine any affected product that remains unused. Immediately cease all further distribution and use of the affected products. Customers should return the product to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Please share this information with appropriate personnel, down to the user level, within your organization or with any organization where the affected devices may have been transferred. Physicians should practice their standard patient monitoring following the procedure for an early identification of any complications to mitigate their severity. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.
Quantity in commerce
106,033
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Product codes and lots
Lots: 14212811 14215685 14212247 14212812 14212814 14215682 14361413 14361416 14285326 14285328 14285330 14297289 14361412 14445407 14444454 14477949 14477950 UDI: (01)00827002474994(17)261013(10)14212811 (01)00827002474994(17)261008(10)14215685 (01)00827002474987(17)261011(10)14212247 (01)00827002474987(17)261008(10)14212812 (01)00827002474994(17)261008(10)14212814 (01)00827002474994(17)261008(10)14215682 (01)00827002474994(17)261219(10)14361413 (01)00827002474994(17)261220(10)14361416 (01)00827002474987(17)261110(10)14285326 (01)00827002474987(17)261110(10)14285328 (01)00827002474994(17)261110(10)14285330 (01)00827002474994(17)261123(10)14297289 (01)00827002474987(17)261220(10)14361412 (01)00827002474987(17)270207(10)14445407 (01)00827002474994(17)270207(10)14444454 (01)00827002474987(17)270217(10)14477949 (01)00827002474994(17)270216(10)14477950

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1240-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1240-2022Class IIOngoingVoluntary: Firm initiated
Z-1241-2022Class IIOngoingVoluntary: Firm initiated
Z-1242-2022Class IIOngoingVoluntary: Firm initiated