Z-1290-2022Class II

Arrow Three-Lumen CVC

Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2022-05-05 / Terminated / root cause: labeling and UDI

ARROW INTERNATIONAL Inc. began a recall of Arrow Three-Lumen CVC on . FDA classified it as Class II and gives the reason as "Gravity flow rates on the affected product lidstock are incorrect". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1290-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Arrow International LLC (subsidiary of Teleflex Incorporated)
Product
Arrow Three-Lumen CVC
Product code
FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Reason for recall
Gravity flow rates on the affected product lidstock are incorrect.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
A recall notification was issued to the consignee on 05/05/2022 via FedEx 2-day mail. The letter instructs the consignee to check the inventory, place a copy of the letter with all affected product and return the response form if the consignee is a medical facility. If the consignee is a distributor, the letter instructs the consignee to provide the notice to their customers, check the inventory, return the completed response form. An updated notification was sent on about 08/09/2022, via FedEx. The updated notification provided an additional description of the label error. Product scope remains the same.
Quantity in commerce
10,301 units
Distribution
US Nationwide distribution in the states of Florida.
Product codes and lots
Catalog Number: CV-12123-F; Lot Numbers: 71F20A0430 [UDI: (01)20801902122838(17)211214(10)71F20A0430], 71F20A2583 [UDI: (01)20801902122838(17)250103(10)71F20A2583], 71F20B0793 [UDI: (01)20801902122838(17)250115(10)71F20B0793], 71F20B2188 [UDI: (01)20801902122838(17)250204(10)71F20B2188], 71F20C1356 [UDI: (01)10801902136944(17)220630(10)71F20C1356], 71F20D1047 [UDI: (01)20801902122838(17)250402(10)71F20D1047], 71F20E2225 [UDI: (01)20801902122838(17)250531(10)71F20E2225], 71F20G2376 [UDI: (01)20801902122838(17)250701(10)71F20G2376], 71F20K0771 [UDI: (01)20801902122838(17)241130(10)71F20K0771], 71F20K2265 [UDI: (01)20801902122838(17)250228(10)71F20K2265], 71F20L0567 [UDI: (01)20801902122838(17)250228(10)71F20L0567], 71F20M1262 [UDI: (01)20801902122838(17)250331(10)71F20M1262], 71F21B0849 [UDI: (01)10801902179514(17)250430(10)71F21B0849], 71F21B0850 [UDI: (01)20801902122838(17)250430(10)71F21B0850], 71F21B2430 [UDI: (01)20801902122838(17)250430(10)71F21B2430], 71F21C0637 [UDI: (01)20801902122838(17)250331(10)71F21C0637], 71F21C0964 [UDI: (01)10801902170856(17)260214(10)71F21C0964], 71F21D0025 [UDI: (01)20801902122838(17)250430(10)71F21D0025], 71F21E0586 [UDI: (01)20801902122838(17)250430(10)71F21E0586], 71F21E1030 [UDI: (01)20801902122838(17)250430(10)71F21E1030], 71F21E2583 [UDI: (01)20801902122838(17)250430(10)71F21E2583], 71F21F1156 [UDI: (01)20801902122838(17)250430(10)71F21F1156], 71F21J0698 [UDI: (01)20801902122838(17)250430(10)71F21J0698]

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1290-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1290-2022Class IITerminatedVoluntary: Firm initiated