Z-1252-2022Class II

Atellica IM 1300 Analyzer - automated, immunoassay analyzer designed to perform in vitro diagnostic tests on...

Siemens Healthcare Diagnostics Inc. / initiated 2022-04-13 / Open, Classified

Siemens Healthcare Diagnostics, Inc. began a recall of Atellica IM 1300 Analyzer - automated, immunoassay analyzer designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN)... on . FDA classified it as Class II and gives the reason as "Lots incompatible with Test Definition (TDef) Version 1.4, potentially lead to the generation of erroneous total T3 results". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1252-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Atellica IM 1300 Analyzer - automated, immunoassay analyzer designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11066001
Product code
JLW Radioimmunoassay, Thyroid-Stimulating Hormone
Reason for recall
Lots incompatible with Test Definition (TDef) Version 1.4, potentially lead to the generation of erroneous total T3 results
FDA root cause
Under Investigation by firm
Action
Siemens issued an Urgent Medical Device Correction (UMDC) to US affected customers by FedEx beginning on 04/13/2022 and an Urgent Field Safety Notice (UFSN) was issued to all OUS. Letter states reason for recall, health risk and action to take: If your facility has multiple lots of T3 available for use and some kit lots are below lots ending in 244, follow the steps below to ensure that T3 results will be correct for all available lots until all kit lots available for use are kit lots ending in 244 or above. Ensure that all IM modules connected to an Atellica Solution have only 1 lot of T3 reagent (all IM modules have the same lot of T3 reagent packs) loaded onto every IM. For reagent kit lots below lots ending in 244: Rescan the Master Curve Card (MCC) for the T3 reagent lot that is on the Atellica IM before processing with that reagent lot, following instructions in the Atellica Online Help (SMN 11313586, section Scanning IM Master Curves and Assay Test Definitions 2D Barcodes ). Note: If the IM is rebooted while using kit lots below lots ending in 244, the MCC will need to be re-scanned before samples are processed. For reagent lots ending in 244 and above: Rescan the Master Curve Card (MCC) for the T3 reagent kit lot that is on the Atellica IM, following instructions in the Atellica Online Help (SMN 11313586, section Scanning IM Master Curves and Assay Test Definitions 2D Barcodes . Reboot all IM modules connected to Atellica Solution following instructions in the Atellica Online Help (SMN 11313586, section titled Shutting Down and Powering Off Analyzers within the System ) when changing from reagent kit lots below lots ending in 244 to a reagent pack associated with kit lots ending in 244 or above before processing with that reagent kit lot. Please review this letter with your Medical Director. Complete and return the Field Action Effectiveness Check Form attached to this letter within 30 days.
Quantity in commerce
1,566
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Argentina, Austria, Bahamas, Bangladesh, Belgium, Brazil, Chile, China, Columbia, Croatia, Czech¿Republic, Denmark, Ecuador, Egypt, France, Germany, Greece, India, Israel, Italy, Japan, Korea, Latvia, Malaysia, Mexico, Netherlands, Paraguay, Peru, Philippines, Portugal, Romania, Russian¿Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, U.A.E., United Kingdom, Vietnam.
Product codes and lots
UDI: 00630414007960 All lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1252-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1252-2022Class IIOngoingVoluntary: Firm initiated
Z-1253-2022Class IIOngoingVoluntary: Firm initiated