Z-1283-2022Class II

CardioTek EP-TRACER Software V2.x

CardioTek BV / initiated 2022-04-25 / Completed

CardioTek BV began a recall of CardioTek EP-TRACER Software V2.x on . FDA classified it as Class II and gives the reason as "Device did not pass electrical safety testing for adequate insulation". As of , the recall status on record is Completed.

Recall record

Recall number
Z-1283-2022
Classification
Class II
Status
Completed
Initiated
Posted
Recalling firm
CardioTek BV
Product
CardioTek EP-TRACER Software V2.x.
Product code
DQK Computer, Diagnostic, Programmable
Reason for recall
Device did not pass electrical safety testing for adequate insulation.
FDA root cause
Under Investigation by firm
Action
The firm issued a recall notification to consignees on 04/25/2022 via email. On 05/24/2022 the firm sent an updated field safety notice. The consignee can continue to use the device. The firm will contact the customer to schedule an appointment for the rework. The consignee is to pass on the letter to those that need to be made aware of the notice.
Quantity in commerce
146 units
Distribution
Worldwide distribution - US Nationwide distribution in the states of CA, NY, NC and the countries of South Korea, Mexico, Kazakhstan, Netherlands, Spain, Germany, Finland, Poland, Israel, Turkey, Chile, Ecuador, Austria, Croatia, Sweden, United Kingdom, Colombia, Italy, Portugal, Canada, Saudi Arabia, Belgium, Slovenia, Brazil, Hungary, Malaysia, Uzbekistan.
Product codes and lots
EP-TRACER 38 UDI/DI: 04260441455418; Serial Numbers: 2019-1-01, 2019-1-02, 2019-1-03, 2019-1-04, 2019-1-05, 2019-1-06, 2019-2-01, 2019-2-02, 2019-2-03, 2019-2-04, 2019-2-05, 2019-3-01, 2019-3-02, 2019-3-03, 2019-3-04, 2019-3-05, 2019-3-06, 2019-3-07, 2019-3-08, 2019-3-09, 2019-3-10, 2019-4-01, 2019-4-02, 2019-4-03, 2019-4-04, 2019-4-05, 2019-4-06, 2019-4-07, 2019-4-08, 2019-4-09, 2019-4-10, 2019-5-01, 2019-5-02, 2019-5-03, 2019-5-04, 2019-5-05, 2020-1-01, 2020-1-02, 2020-1-03, 2020-1-04, 2020-1-05, 2020-1-06, 2020-1-07, 2020-1-08, 2020-1-09, 2020-1-10, 2020-1-11, 2020-1-12, 2020-1-13, 2020-1-14, 2020-1-15, 2020-2-01, 2020-2-02, 2020-2-03, 2020-2-04, 2020-2-05, 2020-2-06, 2020-2-07, 2020-2-08, 2020-2-09, 2020-2-10, 2020-2-11, 2020-2-13, 2020-2-14, 2020-2-15, 2020-4-01, 2020-4-02, 2020-4-03, 2020-4-04, 2021-1-01, 2021-1-02, 2021-1-03, 2021-1-04, 2021-1-05, 2021-1-06, 2021-1-07, 2021-1-08, 2021-2-01, 2021-2-02, 2021-2-03, 2021-2-04, 2021-2-06, 2021-2-07, 2021-2-08, 2021-2-09, 2021-2-10, 2021-2-11, 2021-2-12, 2021-2-13, 2021-2-14, 2021-2-15, 2021-3-01, 2021-3-02, 2021-3-03, 2021-3-04, 2021-3-05, 2021-3-06, 2021-3-07, 2021-3-08, 2021-3-09, 2021-3-10, 2021-3-11, 2021-3-12, 2021-3-13, 2021-3-14, 2021-3-15, 2021-5-01, 2021-5-02, 2021-5-03, 2021-5-04, 2021-5-05, 2021-5-06, 2022-1-01, 2022-1-03, 2022-1-04, 2022-1-05, 2022-1-06; EP-TRACER 102 UDI/DI: 04260441455234; Serial Numbers: 2019-1-11, 2019-2-11, 2019-2-12, 2019-2-13, 2019-2-14, 2019-2-15, 2019-2-16, 2019-4-21, 2019-4-22, 2019-4-23, 2019-4-24, 2019-4-25, 2019-4-26, 2019-4-27, 2020-1-26, 2020-1-27, 2020-1-28, 2020-1-29, 2020-1-30, 2020-2-26, 2020-2-27, 2020-2-28, 2020-2-29, 2020-2-30, 2021-1-13, 2021-1-14, 2021-2-27, 2022-1-26, 2022-1-27

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1283-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1283-2022Class IICompletedVoluntary: Firm initiated