CardioTek BV began a recall of CardioTek EP-TRACER Software V2.x on . FDA classified it as Class II and gives the reason as "Device did not pass electrical safety testing for adequate insulation". As of , the recall status on record is Completed.
Recall record
- Recall number
- Z-1283-2022
- Classification
- Class II
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- CardioTek BV
- Product
- CardioTek EP-TRACER Software V2.x.
- Product code
- DQK Computer, Diagnostic, Programmable
- Reason for recall
- Device did not pass electrical safety testing for adequate insulation.
- FDA root cause
- Under Investigation by firm
- Action
- The firm issued a recall notification to consignees on 04/25/2022 via email. On 05/24/2022 the firm sent an updated field safety notice. The consignee can continue to use the device. The firm will contact the customer to schedule an appointment for the rework. The consignee is to pass on the letter to those that need to be made aware of the notice.
- Quantity in commerce
- 146 units
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of CA, NY, NC and the countries of South Korea, Mexico, Kazakhstan, Netherlands, Spain, Germany, Finland, Poland, Israel, Turkey, Chile, Ecuador, Austria, Croatia, Sweden, United Kingdom, Colombia, Italy, Portugal, Canada, Saudi Arabia, Belgium, Slovenia, Brazil, Hungary, Malaysia, Uzbekistan.
- Product codes and lots
- EP-TRACER 38 UDI/DI: 04260441455418; Serial Numbers: 2019-1-01, 2019-1-02, 2019-1-03, 2019-1-04, 2019-1-05, 2019-1-06, 2019-2-01, 2019-2-02, 2019-2-03, 2019-2-04, 2019-2-05, 2019-3-01, 2019-3-02, 2019-3-03, 2019-3-04, 2019-3-05, 2019-3-06, 2019-3-07, 2019-3-08, 2019-3-09, 2019-3-10, 2019-4-01, 2019-4-02, 2019-4-03, 2019-4-04, 2019-4-05, 2019-4-06, 2019-4-07, 2019-4-08, 2019-4-09, 2019-4-10, 2019-5-01, 2019-5-02, 2019-5-03, 2019-5-04, 2019-5-05, 2020-1-01, 2020-1-02, 2020-1-03, 2020-1-04, 2020-1-05, 2020-1-06, 2020-1-07, 2020-1-08, 2020-1-09, 2020-1-10, 2020-1-11, 2020-1-12, 2020-1-13, 2020-1-14, 2020-1-15, 2020-2-01, 2020-2-02, 2020-2-03, 2020-2-04, 2020-2-05, 2020-2-06, 2020-2-07, 2020-2-08, 2020-2-09, 2020-2-10, 2020-2-11, 2020-2-13, 2020-2-14, 2020-2-15, 2020-4-01, 2020-4-02, 2020-4-03, 2020-4-04, 2021-1-01, 2021-1-02, 2021-1-03, 2021-1-04, 2021-1-05, 2021-1-06, 2021-1-07, 2021-1-08, 2021-2-01, 2021-2-02, 2021-2-03, 2021-2-04, 2021-2-06, 2021-2-07, 2021-2-08, 2021-2-09, 2021-2-10, 2021-2-11, 2021-2-12, 2021-2-13, 2021-2-14, 2021-2-15, 2021-3-01, 2021-3-02, 2021-3-03, 2021-3-04, 2021-3-05, 2021-3-06, 2021-3-07, 2021-3-08, 2021-3-09, 2021-3-10, 2021-3-11, 2021-3-12, 2021-3-13, 2021-3-14, 2021-3-15, 2021-5-01, 2021-5-02, 2021-5-03, 2021-5-04, 2021-5-05, 2021-5-06, 2022-1-01, 2022-1-03, 2022-1-04, 2022-1-05, 2022-1-06; EP-TRACER 102 UDI/DI: 04260441455234; Serial Numbers: 2019-1-11, 2019-2-11, 2019-2-12, 2019-2-13, 2019-2-14, 2019-2-15, 2019-2-16, 2019-4-21, 2019-4-22, 2019-4-23, 2019-4-24, 2019-4-25, 2019-4-26, 2019-4-27, 2020-1-26, 2020-1-27, 2020-1-28, 2020-1-29, 2020-1-30, 2020-2-26, 2020-2-27, 2020-2-28, 2020-2-29, 2020-2-30, 2021-1-13, 2021-1-14, 2021-2-27, 2022-1-26, 2022-1-27
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1283-2022 | Class II | Completed | Voluntary: Firm initiated |