Z-1246-2022Class II

Artis icono floor-angiography systems developed for single and biplane diagnostic imaging and interventional...

Siemens Medical Solutions USA, Inc. / initiated 2022-04-11 / Open, Classified / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Artis icono floor-angiography systems developed for single and biplane diagnostic imaging and interventional procedures SMN: 11327700 on . FDA classified it as Class II and gives the reason as "if the protective earth is interrupted and an additional fault leads to a fault current, the detection of the fault current may not be ensured may cause an electric shock if a person is in contact with metallic parts of...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1246-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Artis icono floor-angiography systems developed for single and biplane diagnostic imaging and interventional procedures SMN: 11327700
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
if the protective earth is interrupted and an additional fault leads to a fault current, the detection of the fault current may not be ensured may cause an electric shock if a person is in contact with metallic parts of the Carm
Root cause tag
design
FDA root cause
Device Design
Action
Siemens Medical Solutions USA, Inc. has initiated a Customer Safety Advisory Notice to all affected customers via AX027/21/S on 4/11/22. Siemens will conduct an inspection of the protective earth via Update Instruction AX026/21/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Acknowledge Receipt of this Safety Advisory Notice Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein.
Quantity in commerce
13 units
Distribution
US Nationwide distribution.
Product codes and lots
UDI-DI: 04056869149325 Serial Number: 170026 170305 170307 170309 170310 170023 170304 170025 170051 170312 170027 170303 170053

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1246-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1245-2022Class IIOngoingVoluntary: Firm initiated
Z-1246-2022Class IIOngoingVoluntary: Firm initiated