ABACUS TPN (Total Parenteral Nutrition) Calculation software, Product codes: 8300-0167 (Abacus V3.1 CE)...
Baxter Healthcare Corporation began a recall of ABACUS TPN (Total Parenteral Nutrition) Calculation software, Product codes: 8300-0167 (Abacus V3.1 CE), 8300-0168 (Abacus V3.1 SE), 8300-0169 (Abacus V3.1... on . FDA classified it as Class I and gives the reason as "There is a potential risk of medication error when using specific Abacus Order Entry & Calculation software". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1376-2022
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- ABACUS TPN (Total Parenteral Nutrition) Calculation software, Product codes: 8300-0167 (Abacus V3.1 CE), 8300-0168 (Abacus V3.1 SE), 8300-0169 (Abacus V3.1 ME), 8300-0191 (Abacus V3.2 CE), 8300-0192 (Abacus V3.2 SE), 8300-0193 (Abacus V3.2 ME), 8300-3391 (Abacus V3.3 CE), 8300-3392 (Abacus V3.3 SE), and 8300-3393 (Abacus V3.3 ME). Commonly used with ExactaMix automated compounding devices (ExactaMix 1200 and ExactaMix 2400).
- Product code
- NEP System/Device, Pharmacy Compounding
- Reason for recall
- There is a potential risk of medication error when using specific Abacus Order Entry & Calculation software.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Baxter sent an Urgent Medical Device Correction notice on 06/22/2022 by letter. The notice explained the problem and the hazard and recommended the following: customers refrain from making updates to bag label templates and to contact Baxter Technical Support if updates to bag label templates are needed. Please continue to follow the instructions in the Abacus User and Abacus Configuration Guides when creating an order, and have a pharmacist thoroughly review all order outputs including the printed label for accuracy and safety. As mentioned above, please ensure your facility processes include pharmacy checks during the compounding process as well as a nursing check. Baxter will communicate to customers when the software upgrade becomes available. At that time additional instructions to obtain the software upgrade will be provided.
- Quantity in commerce
- 9 units
- Distribution
- United States (including Puerto Rico), Argentina, Bermuda, Chile, Colombia, Dominican Republic, Panama, Canada, Vietnam, Hong Kong, Singapore, and EMEA
- Product codes and lots
- Product Code 8300-3391, UDI (01)05413765577345(10)3.3.2.1, Release date 20-Sep-2019; Product Code 8300-3392, UDI (01)05413765577352(10)3.3.2.1, Release date 20-Sep-2019; Product Code 8300-3393, UDI (01)05413765577369(10)3.3.2.1, Release date 20-Sep-2019; Product Codes: 8300-0167, V3.1 CE, Release date 20-May-2013, 8300-0168, V3.1 SE, Release date 20-May-2013, 8300-0169, V3.1 ME, Release date 20-May-2013, 8300-0191, V3.2 CE, Release date 17-Dec-2015, 8300-0192, V3.2 SE, Release date 17-Dec-2015, 8300-0193, V3.2 ME, Release date 17-Dec-2015, No UDIs
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1376-2022 | Class I | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY