Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1274-2022Class II

Colonovideoscope Model CF-H180AL

OLYMPUS Corporation of the Americas / initiated 2022-04-22 / Terminated / root cause: manufacturing process

Olympus Corporation of the Americas began a recall of Colonovideoscope Model CF-H180AL on . FDA classified it as Class II and gives the reason as "A (Camera unit) CCD intended for a different model of colonoscope may have been installed in the CF-H180AL, viewing angle would be narrower than a normal CF-H180AL and the use of EndoTherapy devices may stick out...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1274-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
OLYMPUS Corporation of the Americas
Product
Colonovideoscope Model CF-H180AL
Product code
FDF Colonoscope And Accessories, Flexible/Rigid
Reason for recall
A (Camera unit) CCD intended for a different model of colonoscope may have been installed in the CF-H180AL, viewing angle would be narrower than a normal CF-H180AL and the use of EndoTherapy devices may stick out (approximately 1.4 mm) further than normal protrusion,could lead to patient injury; specifically, mucosal damage that could result in hemorrhage or bleeding
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Olympus issued letter via e-mail on April 22, 2022. Letter states reason for recall, health risk and action to take: 1. Identify the affected model and serial number in the subject of this letter. 2. An Olympus representative will contact you to arrange for the return of your affected device. Olympus will repair your unit free of charge. 3. Olympus will provide a service loaner device in an effort to not disrupt patient care. Contact: 647-999-3203 or at [email protected]
Quantity in commerce
1 unit
Distribution
US Nationwide distribution in the state of KY.
Product codes and lots
UDI-DI: 04953170339615 Serial number: 2501131

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1274-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1274-2022Class IITerminatedVoluntary: Firm initiated