Colonovideoscope Model CF-H180AL
Olympus Corporation of the Americas began a recall of Colonovideoscope Model CF-H180AL on . FDA classified it as Class II and gives the reason as "A (Camera unit) CCD intended for a different model of colonoscope may have been installed in the CF-H180AL, viewing angle would be narrower than a normal CF-H180AL and the use of EndoTherapy devices may stick out...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1274-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- Colonovideoscope Model CF-H180AL
- Product code
- FDF Colonoscope And Accessories, Flexible/Rigid
- Reason for recall
- A (Camera unit) CCD intended for a different model of colonoscope may have been installed in the CF-H180AL, viewing angle would be narrower than a normal CF-H180AL and the use of EndoTherapy devices may stick out (approximately 1.4 mm) further than normal protrusion,could lead to patient injury; specifically, mucosal damage that could result in hemorrhage or bleeding
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Olympus issued letter via e-mail on April 22, 2022. Letter states reason for recall, health risk and action to take: 1. Identify the affected model and serial number in the subject of this letter. 2. An Olympus representative will contact you to arrange for the return of your affected device. Olympus will repair your unit free of charge. 3. Olympus will provide a service loaner device in an effort to not disrupt patient care. Contact: 647-999-3203 or at [email protected]
- Quantity in commerce
- 1 unit
- Distribution
- US Nationwide distribution in the state of KY.
- Product codes and lots
- UDI-DI: 04953170339615 Serial number: 2501131