Z-1257-2022Class II
DJO EMPOWR 3D Knee (Right) 10mm Tibial Insert, REF: 342-10-708
Encore Medical, LP / initiated 2022-05-05 / Open, Classified / root cause: labeling and UDI
Encore Medical, LP began a recall of DJO EMPOWR 3D Knee (Right) 10mm Tibial Insert, REF: 342-10-708 on . FDA classified it as Class II and gives the reason as "A right, size 8, knee tibial insert package may include a mispackaged, left ,size 4, knee tibial insert. If implanted, a risk may include leaving debris in the joint space". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1257-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Encore Medical, LP
- Product
- DJO EMPOWR 3D Knee (Right) 10mm Tibial Insert, REF: 342-10-708
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- A right, size 8, knee tibial insert package may include a mispackaged, left ,size 4, knee tibial insert. If implanted, a risk may include leaving debris in the joint space.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- On 05/05/2022, recall notices were emailed to DJO Surgical Agents by Enovis. Notices state that DJO Surgical requires the following actions: 1) Pass on the recall notice within your organization or to any organization where the potentially affected product has been transferred. 2) Contact Customer Service at 1-800-456-8696 to place a replacement order and to receive an RMA number to return affected devices. 3) Customers with questions or issues can contact Customer Service.
- Quantity in commerce
- 20
- Distribution
- U.S. Nationwide distribution in the states of TN, MN, IL, IN, SC, MS, CA, KS.
- Product codes and lots
- Lot number 078T1141
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1257-2022 | Class II | Ongoing | Voluntary: Firm initiated |