Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1281-2022Class II

BVI Wet -Field Eraser, 18GA Blunt Tip, 45Deg.-Intended to remove tissue and control bleeding by the use of...

Becton Dickinson & Co / initiated 2022-04-26 / Open, Classified / root cause: labeling and UDI

Beaver Visitec International, Inc. began a recall of BVI Wet -Field Eraser, 18GA Blunt Tip, 45Deg.-Intended to remove tissue and control bleeding by the use of high frequency electrical current Part Number: 221251 on . FDA classified it as Class II and gives the reason as "Boxes labeled as Wet-Field Eraser, 18GA Blunt Tip, 45Deg., were shipped containing incorrect product a Wet-Field Eraser, 18GA Bevel Tip, Straight, Part Number: 221250". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1281-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Becton Dickinson & Co
Product
BVI Wet -Field Eraser, 18GA Blunt Tip, 45Deg.-Intended to remove tissue and control bleeding by the use of high frequency electrical current Part Number: 221251
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Boxes labeled as Wet-Field Eraser, 18GA Blunt Tip, 45Deg., were shipped containing incorrect product a Wet-Field Eraser, 18GA Bevel Tip, Straight, Part Number: 221250
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
BVI issued Urgent Medical Device Letter dated 4/26/22 via email stating reason for recall, health risk and action to take: 1. Immediately examine your inventory and quarantine product from these lots that are subject to recall. If you have further distributed this product, please notify your customers by initiating a sub-recall. Consider all potential users of this product in your user supply chain. You are encouraged to use a copy of this recall notification letter when contacting your customers. 2. Complete the enclosed Business Response Form (BRF) and indicate whether you have affected product to return. 3. Please enter the following information on the enclosed BRF: part number(s), lot number(s) and quantity(ies) of device(s) that are being returned. 4. Return ALL quarantined product from the affected lot to our company utilizing the Return Merchandise Authorization (RMA) sticker label that is included with this letter. You may use the prepaid BVI FedEx number for your return. If you have any specific product replacement questions you may contact our Customer Service Team at: [email protected] or call 866-906-8080 from 8 A.M. to 6 P.M. EST (Monday-Friday)
Quantity in commerce
660 units
Distribution
US Nationwide distribution in the states of AZ, CA, CT, DC, FL, GA, IN, MA, MN, NV, SC, TX, WA.
Product codes and lots
UDI-DI: 30886158010488 Lot Number: 6045693

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1281-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1281-2022Class IIOngoingVoluntary: Firm initiated