Z-1336-2022Class II
MassCheck Amino Acid, Acylcarnitines Dried Blood Spot Control Level I
Chromsystems Instruments & Chemicals GmbH / initiated 2022-05-19 / Terminated / root cause: labeling and UDI
Chromsystems Instruments & Chemicals GmbH began a recall of MassCheck Amino Acid, Acylcarnitines Dried Blood Spot Control Level I on . FDA classified it as Class II and gives the reason as "Too low concentration of glycine was detected in the dried blood controls of the batch 2821 of the products". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1336-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Chromsystems Instruments & Chemicals GmbH
- Product
- MassCheck Amino Acid, Acylcarnitines Dried Blood Spot Control Level I
- Product code
- JJY Multi-Analyte Controls, All Kinds (Assayed)
- Reason for recall
- Too low concentration of glycine was detected in the dried blood controls of the batch 2821 of the products.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- The firm issued a recall notification on 05/19/2022 via email. The notification mentions if the consignee has used the affected product, check whether they have determined glycine, replace the old package insert with the updated package insert with release date 18.05.2022, forward a copy of the letter to another laboratory if the affected product was given to them, and submit the response form by 01.06.2022. If users determined glycine, they are asked to clarify with the attending physician or laboratory director, if necessary, whether a back-check of the values for glycine is required.
- Quantity in commerce
- 1330 units
- Distribution
- US: NY, OH, CT.
- Product codes and lots
- Model: 0192; Lot Number: 2821; UDI/DI: 04250317502841
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1336-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1337-2022 | Class II | Terminated | Voluntary: Firm initiated |